Chapter 3
Maintenance
3.
Depress the circuit breaker to reset it.
4.
Reconnect the AC power cable.
5.
Turn on the chassis.
If the circuit breaker trips again, complete the following steps:
1.
Turn off the chassis.
2.
Disconnect the AC power cable.
3.
Remove all modules from the chassis.
4.
Complete the procedure described in the
section of Chapter 2,
Installation and Configuration
. If the power
switch LED is not a steady green, contact National Instruments.
5.
Verify that the NI PXIe-1062Q chassis can meet the power
requirements of your CompactPCI or PXI Express modules.
Overloading the chassis can cause the breaker to trip. Refer to
Appendix A,
6.
The over-current condition that caused the circuit breaker to trip
may be due to a faulty CompactPCI or PXI Express module. Refer to
the documentation supplied with the modules for troubleshooting
information.
Replacing the Modular Power Supply
This section describes how to remove, configure, and install the
AC power supply shuttle in the NI PXIe-1062Q chassis.
Cautions
Disconnect the power cable prior to replacing the power supply.
Do not attempt to use a power supply shuttle from another chassis such as the NI PXI-1042.
Doing so may damage your chassis and the power supply shuttle.
Before connecting the power supply shuttle to a power source, read this
section and the
Read Me First: Safety and Electromagnetic Compatibility
document included with the kit.
Removal
The NI PXIe-1062Q AC power supply is a replacement part for the
NI PXIe-1062Q AC chassis. Before attempting to replace the power supply
shuttle, verify that there is adequate clearance behind the chassis.
Disconnect the power cable from the power supply shuttle on the back of
the chassis. Identify the eight mounting screws for the NI PXIe-1062Q that