1. Introduction  

Thank  you  for  choosing  Champion  F-33  Series  centrifuge  from  Ample 
Scientific. Champion F-33 Series features: 

 

Unique vortex ventilation system keeps chamber temperature  cool 
and preserves sample integrity 

 

Wide-mouth  design  for  convenient  access  to  samples  while 
facilitating centrifuge cleaning and maintenance 

 

Temperature controlled motor to extend centrifuge life expectancy  

 

Well balanced system minimizing vibration and maximize operation 
safety 

2. Product Structure 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 

     

  3. Installation and Setup 

1.

 

Inspect centrifuge for any damages. Contact distributor for 
questions regarding damages. 

2.

 

Place centrifuge on steady and leveled surface. The centrifuge 
should be kept at least 300mm away from surrounding objects in a 
well-ventilated environment. F-33 Series is for indoor use only. 

3.

 

Lid can be unlocked by rotating the ‘Safety Lock’ clockwise. Hold 
‘Safety Lock’ in the unlocked position to open lid. 

4.

 

Make sure all eight of 15ml rotor buckets are installed in rotor. 
Insert provided semi-circular cushions into buckets. Remove any 
loose pieces or packing material in the centrifuge chamber. 

5.

 

Turn the Power Switch on. The Power Switch is located on the back 
of the centrifuge.   

6.

 

Set ‘Timer’ for 1 minute, and ‘Speed’ to highest setting. Press ‘On’ to 
start operation.  

7.

 

 Make sure there is little to no vibration and the whirring sound 
produced is smooth. If excessive vibration or abnormal sound 
occurs, stop the centrifuge and contact distributor immediately 

8.

 

Your centrifuge is now ready to be used. Please use enclosed 
‘Customer Service Card’ to register your machine. Upon registration, 
you will receive a unique Product Warranty Card as your 
certification of authenticity. Customers who failed to register their 
purchased product may be ineligible for manufacturer’s limited 1-
year warranty.  

4. Operation  

1.

 

Turn on centrifuge using power switch on the back. 

2.

 

Open centrifuge lid. Make sure there are no particles or liquid in 
centrifuge chamber.  

3.

 

Make sure all eight of the 15ml rotor buckets are installed into rotor, 
even if you plan to spin less than 8 tubes. 

Do not use buckets from 

another manufacturer or substitute buckets from one centrifuge 
to another*

1

.

  

Figure 1 - Product External Structure 

Safety Lock 

Clear Viewing Port 

Pg | 1 

Pg | 2 

Speed 
Control/Display 

Time 
Control/Display 

Power Switch - on the Back of 
Centrifuge 

 

[On] – Starts 
Operation 

[Off] – Stops 
Operation 