Chapter 6 Safety and Regulatory
Part Number: 0160-6411, Rev A, 02/2019
Page 225 of 314
Software Release: cOS 5.00
6.7 Regulatory Compliance Safety Features
The
CATALYS®
System has been designed to comply with 21 CFR subchapter J as administrated
by the Center for Devices and Radiological Health of the Food and Drug Administration (FDA).
The following compliance safety features are included:
Key Lock Switch
The system can be activated only with the proper key to operate the master key switch. The key
cannot be removed in the
(on) position, and the system will operate only with the key in
place. When treatments are complete, always remove and secure the key to prevent
unauthorized use of the system.
Laser Emission Indicator
The
(laser emission) indicator is displayed on the touchscreen control panel at all times the
laser is energized to warn the user that the system is capable of emitting laser energy.
Door Interlock
A door interlock may be used in conjunction with a remote switch to disable laser and OCT
emission in case of certain external events (e.g., the opening of a treatment room door). A
remote switch or interlock can be wired to the door interlock plug and connected to the system
interlock receptacle on the system rear panel. If a remote switch is used, the system can deliver
laser energy only when the remote switch is closed. Breaking the connection by opening the
switch (door) or removing the plug disables laser and OCT emission, and the system displays a
clearable “Door Interlock” error message on the control panel.
Emergency Stop
The system has an emergency laser stop button that, when pressed, rapidly disables all laser
emission, disables the Mobile Patient Bed or patient chair, and disables the
LIQUID OPTICS™
Interface vacuum system.
Protective Housing
The
CATALYS®
System has a protective housing that prevents unintended human access to laser
radiation. This housing is to be opened only by trained
OPTIMEDICA
personnel.
Safety Interlocks
The protective housing is not designed to be removed by anyone other than an
OPTIMEDICA
-
certified technician. Therefore, the system does not have, and is not required to have, any safety
interlock within the meaning of US FDA 21 CFR, Section 1040, or European EN 60825-1.
However, the protective housing cannot be opened without use of special tools.
Safety Shutter
The
CATALYS®
System includes a safety shutter, which prevents any laser radiation from exiting
the system. The safety shutter is activated when the system is off, during the self-test at turn-on,
in STANDBY mode, or when the safety monitor detects a fault.
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...