Chapter 4 CATALYS® System Procedure
Part Number: 0160-6411, Rev A, 02/2019
Page 175 of 314
Software Release: cOS 5.00
4.2 Intraoperative Instructions
Installing the LIQUID OPTICS™ Interface Components
Prior to seating the patient, install the
LIQUID OPTICS™
Interface components (i.e., disposable
lens and fluid reservoir) on the system.
CAUTION
Inspect all
LIQUID OPTICS™
Interface packaging prior to use. Do not use if the packaging is
damaged or if the device has been dropped, and do not use after the expiration date on the
packaging. Keep the
LIQUID OPTICS™
Interface disposable package sealed and unopened until
immediately prior to use.
WARNING
The
OPTIMEDICA
CATALYS®
System is designed to work with the
OPTIMEDICA
LIQUID OPTICS™
Interface. The
OPTIMEDICA
LIQUID OPTICS™
Interface is the only patient interface approved for
use with the
CATALYS®
System. Do not use any other patient interface device with the
CATALYS®
System. Do not modify or alter the
LIQUID OPTICS™
Interface used with the
CATALYS®
System.
Figure 4.14 Location of
LIQUID OPTICS™
Interface Component Installation
LIQUID OPTICS™
Interface
Circular Label Reader
LIQUID OPTICS™
Interface
Vacuum Connections
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...