Chapter 4 CATALYS® System Procedure
Part Number: 0160-6411, Rev A, 02/2019
Page 163 of 314
Software Release: cOS 5.00
Table 4.1 User-adjustable Capsulotomy Parameters
Table 4.2 Capsulotomy Parameters and Safety Margins
Feature
Default
Range
Step Size
Units
Pattern
Circle
N/A
N/A
N/A
Depth
600
200-1000
200
µm
Diameter
N/A
2.0-8.0
0.1
mm
Horizontal Spot Spacing
5
3-10
1
µm
Vertical Spot Spacing
10
5-50
5
µm
Laser Pulse Energy
4
1-10
0.5
µJ
Safety Margins
Feature
Value
Units
Iris
500
µm
Corneal
500
µm