User Manual for AML Oceanographic’s MVP

Communications 

Factory Default Settings 

MVP•X have the following settings: 

 

RS-485 

 

19200 BAUD rate 

 

Data on power up 

 

Sampling rate of 25 Hz 

Interrupting the Monitor Command

 

MVP•X are set to 

not

 check the RX line and to output data continuously on power up. To interrupt 

that data flow to, for example, alter the default factory settings, you must interrupt the monitor 
command by performing the following steps: 

 

Open a terminal emulation program such as HyperTerminal. Ensure the serial port has 

been selected in the program and that the appropriate baud rate has been selected. Note 
that MVP•X are set to 19,200 baud at the factory. 

 

Connect the instrument to the computer using the data/power cable 

 

Unplug the data/power cable from the instrument to turn the instrument off. 

 

Hold down the < ENTER > key. 

 

Plug the data/power cable into the instrument to turn the instrument on. 

 

Release the <ENTER> key once the prompt ‘>’ is displayed. 

MVP•X will now accept command line input. See the Appendix for a list of advanced commands. 

Returning MVP•X to Factory Default Settings for Use with Rolls-Royce MVPs

 

As MVP•X is set to ignore character input and output data on power up, it is necessary to interrupt 
the data flow. See “Interrupting the Monitor Command.” Once that is complete, perform the 
following steps: 

 

Open a terminal emulation program such as HyperTerminal. Ensure an RS-485 serial port 
has been selected in the program.  By default, MVP•X should be set to 19,200 baud.  The 
terminal emulation program must be configured to the appropriate baud rate. 

 

Connect the instrument to the computer using a data/power cable. Using the terminal 

emulation program, issue the following commands to the instrument: 

o

 

SET S C

 

 

 

(set sample rate continuous at 25 Hz) 

o

 

SET DETECT 06

 

 

(set fixed BAUD rate of 19200) 

o

 

SET EMU BROOKE

 

 

(set P,SV,C and T output order) 

o

 

SET RX OFF

   

 

(disables reception of characters at next power 

up) 

o

 

SET STARTUP NOHEADER

 

(disables the power up header information) 

o

 

SET STARTUP MONITOR

 

(enables data output on power up) 

o

 

SET SCAN  NOTIME

   

(removes time column from output) 

o

 

SET SCAN  NODATE

   

(removes date column from output) 

o

 

SET SCAN  NOSALINITY 

(removes salinity column from output) 

o

 

SET SCAN  NOBATT   

(removes battery voltage from output) 

                          

Note: Details on the 

SET DETECT

 command can be found in the Appendix. 

Содержание MVP-X

Страница 1: ...MVP X User Manual Version 1 02 4 MAR 2015...

Страница 2: ...Revision History Revision Date Changes Name Version 1 0 27 Oct 2014 Created manual DO Version 1 01 4 Mar 2015 Formatting Edits JZ Version 1 02 22 June 2015 Updating technical specifications JZ...

Страница 3: ...Factory Default Settings 7 Interrupting the Monitor Command 7 Returning MVP X to Factory Default Settings for Use with Rolls Royce MVPs 7 Output Formats 8 Default Output Format 8 Support 9 Troublesho...

Страница 4: ...Calibration data for all sensors is available from the instrument and calibration certificates can be printed from AML Oceanographic s SeaCast software whenever the instrument is connected Logged data...

Страница 5: ...he instrument Xchange sensor manuals C Xchange SV Xchange P Xchange and T Xchange providing overviews on how to install and maintain each of the Xchange sensors If you are configuring an instrument fo...

Страница 6: ...on Inspect for damage o Check the cable for slices or gouges o Check the connector sockets for corrosion dirt and salt deposits o Check the pressure case for dents and scrapes o Check the sensors for...

Страница 7: ...and 6000dBar Pre Deployment Procedures Upon Receipt o Use the Shipping and Receiving instructions to verify the condition of the instrument o Verify that all sensor calibrations are valid for the dur...

Страница 8: ...ore each use o Check for proper installation of all Xchange sensors o Check for nicks and cuts on the cable After each use o Clean and rinse the instrument using fresh water o Dry the instrument compl...

Страница 9: ...MVP X to Factory Default Settings for Use with Rolls Royce MVPs As MVP X is set to ignore character input and output data on power up it is necessary to interrupt the data flow See Interrupting the M...

Страница 10: ...t for each Xchange sensor Sensor Units Default Format SV Xchange m s 1234 567 Conductivity Xchange mS cm 12 346 Pressure Xchange dBar 1234 56 Temperature Xchange C 12 345 Example CTD output 0000 61 33...

Страница 11: ...instrument specifically set to one baud rate last time If so use that baud rate to resume communications o Was the instrument set to only RS 232 or only RS 485 last time If so resume communications in...

Страница 12: ...iated with each individual sensor and Chen and Millero s equation1 sound velocity calculated from CTD values will fall within approximately 0 4 m s of the actual value 95 of the time It is not uncommo...

Страница 13: ...s prospects and other members of our community to manage their interactions with AML My AML Oceanographic will allow you to View and manage your assets instruments and sensors Consult instrument diagn...

Страница 14: ...AUTO DISPLAY DETECT Displays the baud rate detection settings SET DETECT a b Sets the baud rate detection a sets the number of autobaud detection attempts before the instrument reverts to the default...

Страница 15: ...ing DISPLAY SCAN C X T X and P X SET SCAN SV Turns the calculated sound velocity channel on allowing it to be present in instrument output scans Current salinity display status is viewable using DISPL...

Страница 16: ...s barometric offset P X DIS TIME Displays current time Time format is hh mm ss ss SET TIME hh mm ss ss Sets instrument time using 24 hour clock in format hh mm ss ss DETECT Checks each slot in logger...

Страница 17: ...ms at 1 m s flow SV Xchange 1375 to 1625 m s 0 025 m s 0 006 m s 0 001 m s 47 s Secondary Xchange Sensors Type Range Accuracy Precision Resolution Response Time Temperature Xchange 2 C to 32 C 0 005 C...

Страница 18: ...6 VDC Current Draw at 12VDC Primary Sensor Secondary Sensor Secondary Sensor Current Draw mA 119 C X 122 C X P X 129 C X T X 138 C X P X T X 145 SV X 120 SV X P X 127 SV X T X 135 SV X P X T X 145 T X...

Страница 19: ...up Included Items MVP X instrument One primary sensor blanking plug Two secondary sensor blanking plugs USB stick with manuals and documentation Housing Status Type Depth Rating Diameter Length Weigh...

Страница 20: ...05 MVP X 500 dBar SVPT Configuration PDC MVPX SVPT 10 MVP X 1000 dBar SVPT Configuration PDC MVPX SVPT 40 MVP X 4000 dBar SVPT Configuration PDC MVPX SVPT 60 MVP X 6000 dBar SVPT Configuration Sensor...

Страница 21: ...hic s MVP X 19 XCH PRS 2000 P Xchange 2000 dBar XCH PRS 4000 P Xchange 4000 dBar XCH PRS 5000 P Xchange 5000 dBar XCH PRS 6000 P Xchange 6000 dBar XCH PRS 6000 T065 P Xchange 6000 dBar Extended temper...

Страница 22: ...oes not apply if the instrument has been damaged by accident or misuse and is void if repairs or modifications are made by any other than authorized personnel This warranty is the only warranty given...

Страница 23: ...User Manual for AML Oceanographic s MVP X 21 Technical Overview Drawings...

Страница 24: ...X CTD 10 PKG MVPX CTD 40 PKG MVPX CTD 60 PKG MVPX SVTP 05 PKG MVPX SVTP 10 PKG MVPX SVTP 40 PKG MVPX SVTP 60 PKG MVPX CTD XX THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF AML OCEAN...

Страница 25: ...White 2 3 1 Pin 1 3 5 2 6 PMCIL 6 FS 4 2071 MALAVIEW AVE W SIDNEY B C CANADA V8L 5X6 PH 250 656 0771 FAX 250 655 3655 www AMLoceanographic com DATE SLDRW NO 11SEP2014 TITLE SYSTEM MC3 PMCIL 6 FS TO RS...

Отзывы: