Installation Procedure
1. Mount analyzer in a shelter if convenient, at a convenient eye level, not too close to the
right wall (so as not to interfere with the gas connections).
2. Make the gas connections
a. Connect the sample line with ¼” ss tubing.
b. Make sure that the sample gas line has a pressure no greater than 150 psig.
c. If so, use a pressure reducing regulator and adjust the inlet pressure to 25 psig.
d. Connect the exhaust to a line that goes outside the building, or to a suitable purge
system. Make sure it runs downhill so no condensation from a wet sample forms
ice or water and blocks it up in it in winter.
e. Turn the sample valve to SAMPLE.
f. Adjust sample flow to 1 SCFH with the flow control valve.
We recommend that you use an “Analyzer Guardian” (a liquid rejection device) on the sample line if there is
any possibility of water or entrained liquids in the sample.
Don’t allow liquid water to enter the analyzer!
This may mean that you have to cool the sample and drop out water before it gets into the analyzer
.