PN 9000-280-VE, Rev B
Installing the User Interface/Communications Options
The WDG-V UOP Analyzer has two available interface options that can be used
to communicate with the analyzer: The AMEVision Display User Interface and
the PC Configurator Software. These interface options cannot be connected to
the analyzer at the same time.
Installing the PC Configurator Software (Typical
Communication Option)
The WDG-V PC Configurator Software comes standard with the analyzer and is
used to configure, calibrate, and monitor a single analyzer. The WDG-V UOP is
connected to the PC using a two-wire MODBUS RTU interface.
If you are using the AMEVision Display User Interface to communicate
with the WDG-V UOP Analyzer, skip this section and continue with
“Installing the AMEVision Display User Interface (Secondary Commu-
nication Option)” later in this chapter.
Install the PC Configurator Software on a PC, as instructed in the
“WDG-V PC Configurator Software Manual” (AMETEK PN 9000-231 VE),
and then continue with setting the analyzer address, below.
Setting the Analyzer Address Using the Dip Switch (to
Enable Communications Between the Analyzer and the PC
Configurator Software)
The address on the WDG-V UOP analyzer is set using the dip switch on the front
of the
Sensor
board. When connecting a single analyzer, the address is “
1
”.
Before continuing,
install the PC Configurator
Software on a PC, as
instructed in the “WDG-V
PC Configurator Software
Manual” (AMETEK PN
9000-231-VE).
The orientation of the
dip switch is upside down.
321 Switch
Address
001
1
1 = ON (UP)
0 = OFF (DOWN)
Figure 2-11.
Close-up of Dip Switch, Sensor
board.
Installation and Start-Up
|
2-27
Содержание WDG-V UOP
Страница 1: ...WDG V UOP Combustion Analyzer ATEX IECEx USER MANUAL PN 9000 280 VE_Rev C...
Страница 40: ...PN 9000 280 VE Rev B Figure 2 7 Conduit Entries 2 18 Thermox WDG V UOP ATEX IECEx Analyzer...
Страница 96: ...PN 9000 280 VE Rev B WDG V UOP Series Analyzer Rating Labels 4 6 Thermox WDG V UOP ATEX IECEx Analyzer...