PN 9000-280-VE, Rev B
Installing the Analyzer
• The analyzer ambient temperature range is -20 °C to 60 °C.
• When installing the exhaust tube, attach it to the back of the analyzer
before mounting the analyzer to the process.
• Mounting dimensions are provided in Figure 2-2.
Do not open a Z-purge, hinged-analyzer enclosure door until you first
verify that the area has been classified as non-hazardous.
The factory-supplied mounting flange comes attached to the UOP
analyzer. Mount the analyzer flange to the process flange as described
below.
1. Thread the 1/4 -18 NPT sample inlet probe into the mounting flange.
Orient the probe so the longer side faces towards the flow of gases; this
prevents large particles from entering the analyzer. This inlet pipe does not
need to be leak tight.
When mounting the analyzer to your process flange (see Step 2), the
mounting flange must also be sealed to your process flange using the
factory-supplied Inconel compressible gasket. Center this gasket on
the raised flange faces to ensure a leak-tight seal. You can use a thin
piece of sheet metal to help you install this gasket.
2. Mount the factory-provided 3" 300# mounting flange to your process
flange. Tighten the nuts onto the mounting studs in a cross pattern to cre-
ate an even seal (orient the flange so the analyzer is in an upright position).
This flange has eight 3/4" - 10 studs, 3-1/2" long that conform to ASTM
A193 B8M. Heavy hex nuts (ASTM A194 8M), 3/4" - 10 threads, and associ-
ated washers are also included.
Mounting dimensions for the hinged style enclosure, including the Z-purge
option, are provided in Figures 2-2 and 2-4.
Installation and Start-Up
|
2-7