Controller / User Interface | 4-15
Set Current Range
Define the range of readings that the 0-to-20 or 4-to-20 mA output will
represent. You can also choose to set a reverse analog output range. For
example, 0 mA can correspond to the high end of the oxygen range, and
20 mA can correspond to the low end of the oxygen range.
The range of readings the current output represents is also a function of
the type of process readings the analog output port tracks (see the “Select
Function” section later in this chapter for the types of readings the current
output can represent).
See the “Set Current Mode” section for help on choosing either a 0-to-
20 mA or 4-to-20 mA scale.
Setting the current range
1. Select an analog output port.
2. From the
Analog Range
key menu, select
Set Current Range
.
You will be prompted to define the new value to correspond to the 20 mA
analog output:
Out # y 20 ma = xx
New Value?
where y is the analog output port, and xx is the present value that the cur-
rent output currently represents (for example, 10 % O
2
). The output port
that appears depends on the analog port you selected.
3. Input the value to correspond to the 20 mA analog output.
You will be prompted to define the value to correspond to the 0 or 4 mA
analog output:
Out # 1 4 ma = xx
New Value?
where
xx
is the value that the current output currently represents.
4. Input the value to correspond to the 4 (or 0) mA analog output.
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...