Calibration | 5-17
Cal Gas Value
This menu option allows you to enter span and zero calibration gas
values to match the calibration gas cylinders you use to calibrate or
verify your system. These values will be used for all automatic, remote,
or manual calibrations and verifications.
•
O
2
span gas - high calibration gas
•
Zero gas - low calibration gas
•
Combustibles span gas (optional)
The O
2
span gas value you enter must be higher than the O
2
zero
gas value.
1. Select
Cal Gas Value
from the Calibrate key menu.
You are prompted to enter the span gas value:
Span Gas = xx.x% O2
New Value?
2. Enter the span gas value, then press the
Enter
key. You can also
include a decimal point (for example, 20.9% O
2
). If you want to keep
the current span gas value unchanged, press the
Enter
key without
entering a value.
You are then prompted to enter the zero gas value:
Zero Gas = xx.xx% O2
New Value?
3. Enter the zero gas value, then press the
Enter
key. You can enter up
to two decimal points (for example, 2.00% O
2
). If you want to keep
the current zero gas value unchanged, press the
Enter
key without
entering a value.
NOTE
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...