5-10 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Remote Calibrate/Remote Verify
To initiate a remote calibration or verification, do the following:
1. Enter the calibration gas values to match the span and zero calibra-
tion gas cylinders you intend to use to calibrate/verify your system
(see the “Cal Gas Values” section later in this chapter for help on how
to do this). These values are referred to in this section as setpoint
values.
2. Select
Initiate Cal
or
Initiate Verify
from the Calibrate key menu,
depending on whether you want to perform a calibration or verifica-
tion.
3. Select
Remote Calibrate
or
Remote Verif
y.
The following then appears on your display (display line 2 will only
appear if you have the combustibles option):
O2 xx.x % Set xx.x %
Cmb y
Press ENTER to Span
4. Press the Enter key once the O
2
reading on the far left of the first
display line stabilizes. The reading on the far right of the first display
line shows the setpoint value you entered using the Cal Gas Values
menu option. This setpoint value should match the cal gas cylinder
value.
•
If your O
2
reading on the far left of the display has stabilized, yet is not
near the setpoint value on the far right of the display, you should check
the cal gas value you entered and ensure that this value matches the cal gas
cylinder for that gas in the RCU.
When performing a primary calibration or the first calibration after
a primary calibration, the reading and the setpoint values often
won’t match. Once the first regular calibration after a primary
calibration is performed, the readings should match.
NOTE
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...