3. APPLICATION AND STANDARD REQUIREMENTS

3.1. General

The KEMA 801A is recommended as an additional decoupling network (ferrite tube clamp) for immunity test 
according to IEC / EN 61000-4-6 when using the clamp injection method. It shall be inserted on all the cables 
between EUT and AE except the cable under test. The KEMA 801A prevents, that the test signal applied to the 
EUT affects other devices, equipment or systems, which are not under test and improves the reproducibility 
of the test results.
The clamp mechanism of the KEMA 801A makes possible a simple insertion in the test setup.
Other application: Prevention of unwanted RF- currents at cables during emission measurements. 

3.2. Test set-up example

The following figure shows a test set-up example for testing in accordance with IEC 61000-4-6 (EN 61000-4-

6). 

Figure 1:   Test set-up with EUT according IEC / EN 61000-4-6 with EM clamp (KEMZ 801), RF 

attenuation clamp (KEMA 801A) and monitoring probe (MD 4070)

AE-side

EUT-side

KEMZ 801

Insulating

Ground plane

6 dB attenuator

Auxiliary

equipment

150 kHz  ...  230 MHz  (1000 MHz)

EM-Koppelzange

KEMZ 801

EUT

RF out

Amp in

< +10 dBm

Amp out

ch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBm

Power meter

NSG 4070

Power

USB

Tuning

StSize

StSize

StSize

Local

Back

Stop

Run

Hold

0

.

1

4

7

8

5

2

3

kHz

dBm

6

9

MHz

dBµV

Hz

V

Enter

Step

1

STO

FRQ

LVL

RCL

Step

2

Step

3

MOD

2nd

Help

RF

ON/OFF

EUT

AE 1

Equipment

under test

KEMA 801A

Ground plane

150 kHz  ...  1000 MHz 

RF Attenuation Clamp

KEMA 801

MD 4070

Power meter

Insulating

AE 2

Insulating

Test generator e.g. Teseq NSG 4070

with built-in RF synthesizer, 

broadband power amplifier and 

power meter

Содержание TESEQ KEMA 801A

Страница 1: ...USER MANUAL RF ATTENUATION CLAMP KEMA 801A...

Страница 2: ...USER MANUAL RF ATTENUATION CLAMP KEMA 801A...

Страница 3: ...1 General 6 2 2 Storage and transport 6 2 3 Unpacking 6 2 4 Scope of delivery and order information 6 3 Application and standard requirements 7 3 1 General 7 3 2 Test set up example 7 4 Verification...

Страница 4: ...4 KEMA 801A...

Страница 5: ...rongly recommended Operate the equipment only in dry surroundings Allow any condensation that occurs to evaporate before putting the instrument into operation Do not exceed the permissible ambient tem...

Страница 6: ...upwards Protect from dampness heat cold and rain Do not throw Do not sit or stand on the instrument and packaging 2 3 Unpacking Is the packaging damaged If YES transportation company Are all the pack...

Страница 7: ...p example The following figure shows a test set up example for testing in accordance with IEC 61000 4 6 EN 61000 4 6 Figure 1 Test set up with EUT according IEC EN 61000 4 6 with EM clamp KEMZ 801 RF...

Страница 8: ...gure 3 Set up for measuring the insertion loss 2x 10 dB attenuator RF OUT 50 RF IN 50 NETWORK ANALYZER USB USB Power Tuning Back 0 1 4 7 8 5 2 3 kHz dBm 6 9 MHz dB V Hz V Enter StSize StSize StSize St...

Страница 9: ...9 85 238526 E02 5 CONSTRUCTION OF THE PRODUCT Figure 4 View to the opened KEMA 801A Ferrite cores spring loaded Mechanical interlock adjustable Figure 5 Side view...

Страница 10: ...79 mm Weight approx 5 5 kg Environmental classification Indoor use only Operating temperature 5 C to 40 C Relative humidity up to 80 Typical insertion loss in 50 system 0 0 10 20 40 30 60 50 0 1 1 10...

Страница 11: ...ries by the user 7 2 Cleaning The cleaning shall be done with dry cloth If a wed cleaning would become necessary make sure that no humidity gets inside the unit and clean the instrument housing with a...

Страница 12: ...rise Shanghai Co Ltd Beijing Branch T 8610 8526 2111 F 8610 8526 2141 chinasales teseq com Germany Teseq GmbH T 49 30 5659 8835 F 49 30 5659 8834 deinfo teseq ametek com Singapore AMETEK Singapore Pte...

Отзывы: