Local and Remote Sensing
TM-F1OP-C1XN-01
2-23
Local and Remote Sensing
Use connections at the rear panel J10 sense connector to configure the
power supply for local or remote sensing of output voltage. See
for a drawing of the sense connector.
Sense Wiring
For lowest noise performance, use shielded pair wiring of 16 to 24 AWG
for remote sense lines. Strip wires 0.26 in. (6.5 mm), insert and tighten the
wire clamp.
WARNING
There is a potential shock hazard at the sense connector when using a
power supply with a rated output greater than 40 V. Select wiring with a
minimum insulation rating equivalent to the maximum output voltage of
the power supply for use as local sense jumpers or for remote sense wires.
For example, select TEW-105, 105 °C, 600 V wiring for use with a 600 V,
2 A model power supply. Ensure that connections at the load end are
shielded to prevent contact with hazardous voltages.
TM-F1OP-C1XN-01.book Page 23 Thursday, October 25, 2007 9:04 PM
Содержание Sorensen XFR 1200 Watt Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ii...
Страница 8: ...This page intentionally left blank iv...
Страница 10: ...This page intentionally left blank vi...
Страница 16: ...xii TM F1OP C1XN 01 book Page xii Thursday October 25 2007 9 04 PM...
Страница 18: ...xiv TM F1OP C1XN 01 book Page xiv Thursday October 25 2007 9 04 PM...
Страница 30: ...1 12 TM F1OP C1XN 01 book Page 12 Thursday October 25 2007 9 04 PM...
Страница 57: ...Local and Remote Sensing TM F1OP C1XN 01 2 27 TM F1OP C1XN 01 book Page 27 Thursday October 25 2007 9 04 PM...
Страница 58: ...2 28 TM F1OP C1XN 01 book Page 28 Thursday October 25 2007 9 04 PM...
Страница 82: ...4 8 TM F1OP C1XN 01 book Page 8 Thursday October 25 2007 9 04 PM...
Страница 94: ...5 12 TM F1OP C1XN 01 book Page 12 Thursday October 25 2007 9 04 PM...