Table 3-2. Remote Power-on Conditions
Condition
Default
Voltage
0 Volts (initial from factory power-on voltage); otherwise,
last value saved by SCPI command.
See CAL:INIT:VOLT
Current
0 Amps (initial from factory power-on current); otherwise,
last value saved by SCPI command.
See CAL:INIT:CURR
Soft Voltage Limit
Model maximum voltage
*
Soft Current Limit
Model maximum current
*
OVP Trip Voltage
Model maximum v10% (initial from factory power-
on OVP); otherwise, last value saved by SCPI command.
See CAL:INIT:VOLT:PROT
Delay
0.5 seconds
Foldback Protection
OFF (non-configurable)
Output
ON
**
See CAL:MOD:POWERON
Hold
OFF
Unmask
NONE
Service Request
Capability
OFF
* User-programmable temporary limit (reverts to power-on defaults after power
cycle or Reset command is issued).
** User-selectable.
Содержание Sorensen SG Series
Страница 2: ......
Страница 4: ...Contents SG Series Programming viii M550129 03 Rev L This page intentionally left blank...
Страница 6: ...Contents SG Series Programming x M550129 03 Rev L This page intentionally left blank...
Страница 8: ...Contents SG Series Programming xii M550129 03 Rev L This page intentionally left blank...
Страница 14: ...Contents SG Series Programming xviii M550129 03 Rev L This page intentionally left blank...
Страница 16: ...Overview SG Series Programming 1 2 M550129 03 Rev L This page intentionally left blank...
Страница 126: ...SCPI Command Operation SG Series Programming 6 60 M550129 03 Rev L This page intentionally left blank...
Страница 143: ...M550129 03 Rev L 8 1 SECTION 8 SCPI STATUS IMPLEMENTATION...
Страница 144: ...SCPI Status Implementation SG Series Programming 8 2 M550129 03 Rev L This page intentionally left blank...
Страница 146: ...This page intentionally left blank...