SG Series Programming
SCPI Command Operation
M550129-03 Rev L
6-51
OUTP:STAT ON
PROG:STAT RUN
Following is a diagram for Example 4.
Figure 6-4. End-of-Sequence Pause for Example 4
EXAMPLE 5
To make a sequence longer than the maximum of 20 steps in an ordinary sequence, there
are either one of two ways or a combination of the two. One way is to use the
SUBCALL
command. The second way is to use the
GOTO
command. The use of the
SUBCALL
command was shown in Examples 3 and 4 above. Now join two 20-step sequences with a
GOTO
command.
First, create a sequence wherein the power supply output will simulate a capacitor charge
curve to a maximum of 10 volts and discharge curve to zero volts.
3 volts for 10 sec
3 volts for 10 sec
5 volts for 10 sec
Ramp up to 5
volts in 10 sec
Ramp down to 3
volts in 10 sec
10 volts for 5 sec
Ramp down to
2 volts in 9 sec
4 volts for 6 sec and
then pause indefinitely
at these last settings.
Содержание Sorensen SG Series
Страница 2: ......
Страница 4: ...Contents SG Series Programming viii M550129 03 Rev L This page intentionally left blank...
Страница 6: ...Contents SG Series Programming x M550129 03 Rev L This page intentionally left blank...
Страница 8: ...Contents SG Series Programming xii M550129 03 Rev L This page intentionally left blank...
Страница 14: ...Contents SG Series Programming xviii M550129 03 Rev L This page intentionally left blank...
Страница 16: ...Overview SG Series Programming 1 2 M550129 03 Rev L This page intentionally left blank...
Страница 126: ...SCPI Command Operation SG Series Programming 6 60 M550129 03 Rev L This page intentionally left blank...
Страница 143: ...M550129 03 Rev L 8 1 SECTION 8 SCPI STATUS IMPLEMENTATION...
Страница 144: ...SCPI Status Implementation SG Series Programming 8 2 M550129 03 Rev L This page intentionally left blank...
Страница 146: ...This page intentionally left blank...