M550004-01 Rev K 

www.programmablepower.com 

 

DHP Series 

DC Power Supplies 

Operation Manual  

Содержание Sorensen DHP Series

Страница 1: ...M550004 01 Rev K www programmablepower com DHP Series DC Power Supplies Operation Manual...

Страница 2: ......

Страница 3: ......

Страница 4: ......

Страница 5: ...fication purposes only Notice of Copyright DHP Series DC Power Supplies Operation Manual 2002 AMETEK Programmable Power Inc All rights reserved Exclusion for Documentation UNLESS SPECIFICALLY AGREED T...

Страница 6: ...This page intentionally left blank ii...

Страница 7: ...and equipment safety Always ensure that facility AC input power is de energized prior to connecting or disconnecting any cable In normal operation the operator does not have access to hazardous volta...

Страница 8: ...This page intentionally left blank iv...

Страница 9: ...nstructions of AMETEK is opened modified or disassembled in any way without AMETEK s consent or is used in combination with items articles or materials not authorized by AMETEK The Buyer may not asser...

Страница 10: ...This page intentionally left blank WA 2...

Страница 11: ..._______________ 3 8 3 4 REMOTE CURRENT PROGRAMMING __________________________________ 3 8 3 5 REMOTE VOLTAGE PROGRAMMING __________________________________ 3 9 3 6 REMOTE SENSING _____________________...

Страница 12: ...uto step Programming Screen 4 1 Figure 4 2 Auto step Programming Menu System 4 2 Figure 4 3 Programming Start Screen 4 3 Figure 4 4 Control Source Menu 4 4 Figure 4 5 Control Source Auto step 4 4 Figu...

Страница 13: ...ude three separate series the 2U High the 3U High and the 6U High The power supplies of each series are constant current constant voltage supplies with an automatic crossover feature The 2U High Serie...

Страница 14: ...nnections not included 3 50 in 45 lbs max 208 230 VAC tested to 190 253 VAC 47 to 63 Hz Single 2 wire plus ground Three 3 wire plus ground optional 0 1 of max output voltage 0 2 of max output current...

Страница 15: ...Temperature Operating Storage Humidity Cooling Agency Approvals excludes 600 volt models 0 5k ohms 0 5k ohms 0 5 10 VDC 0 5 10 VDC Terminals are provided to sense output voltage at point of load Maxi...

Страница 16: ...ions not included 5 25 in 10 5 in 120 lbs max 163 lbs max 208 230 VAC tested to 190 253 VAC 400 VAC tested to 360 440 VAC 480 VAC tested to 432 528 VAC 47 to 63 Hz 3 phase 3 wire plus ground 0 1 of ma...

Страница 17: ...mperature Operating Storage Humidity Cooling Agency Approvals excludes 600 volt models 0 5k ohms 0 5k ohms 0 5 VDC or 0 10 VDC 0 5 VDC or 0 10 VDC Terminals are provided to sense output voltage at poi...

Страница 18: ...50 250 8 12 20 40 60 80 100 120 300 6 6 10 16 33 50 66 83 100 400 5 7 5 12 25 37 50 62 75 600 11 16 22 27 33 Notes Basic model number includes voltage and current from from table e g 600 volt models a...

Страница 19: ...odels respectively Tables 2 3 and 2 4 provide input and output connector descriptions For permanently connected equipment a readily accessible disconnect device shall be incorporated in the fixed wiri...

Страница 20: ...ble 2 3 see Table 2 3 Control Interface 190 253 VAC Std 47 63 Hz Power Source User load s See Table 3 2 for description Table 2 2 5kW to 15kW and 16kW to 30kW Series Input Output Connectors CONNECTOR...

Страница 21: ...t terminal exceed 150 volts positive or negative from chassis potential Table 2 4 Input Connection Descriptions Note Observe the maximum torque specification indicated on the cover INPUT VOLTS PHASE C...

Страница 22: ...ES TYPES TYPES AWG MCM RUW T TW UF FEPW RH RHW RUH THW THWN XHHW USE ZW V MI TA TBS SA AVB SIS FEP FEPB RHH THHN XHHW 14 20 20 25 25 12 25 25 30 30 10 30 35 40 40 8 40 50 55 55 6 55 65 70 75 4 70 85 9...

Страница 23: ...3kW Output Voltage 60 V M550004 01 2 5 AC CONTROL STATUS CONSTANT POWER CONSTANT CURRENT REMOTE CONSTANT VOLTAGE DC CONTROL 3 2 1 MENU LAST SET 6 9 4 7 5 0 8 ENTER IEEE ADDRESS J2 J1 REMOTE PROGRAM RE...

Страница 24: ...3kW Output Voltage 80 V M550004 01 2 6 AC CONTROL STATUS DC CONTROL CONSTANT VOLTAGE CONSTANT CURRENT REMOTE CONSTANT POWER MENU LAST SET 5 4 6 1 2 3 8 0 7 9 ENTER IEEE ADDRESS J2 NEG J3 DC OUTPUT POS...

Страница 25: ...Figure 2 3 Outline Drawing 3 3kW to 15kW Output Voltage 60 V AC CONTROL STATUS REMOTE CONSTANT VOLTAGE CONSTANT CURRENT CONSTANT POWER MENU LAST SET DC CONTROL 5 2 8 0 4 1 7 6 3 9 ENTER M550004 01 2 7...

Страница 26: ...Figure 2 4 Outline Drawing 3 3kW to 15kW Output Voltage 80 V AC CONTROL STATUS CONSTANT CURRENT CONSTANT POWER REMOTE CONSTANT VOLTAGE 6 4 5 MENU LAST SET 9 7 0 8 DC CONTROL 3 1 2 ENTER M550004 01 2 8...

Страница 27: ...e 2 5 Outline Drawing 13 3kW 16kW to 30kW Output Voltage 60 V AC CONTROL STATUS REMOTE LAST SET 0 CONTSTANT VOLTAGE CONTSTANT CURRENT CONTSTANT POWER DC CONTROL MENU 1 4 7 2 5 8 3 6 9 ENTER M550004 01...

Страница 28: ...e 2 6 Outline Drawing 13 3kW 16kW to 30kW Output Voltage 80 V STATUS AC CONTROL REMOTE CONTSTANT VOLTAGE CONTSTANT CURRENT CONTSTANT POWER 9 8 7 LAST SET 0 MENU DC CONTROL 2 1 5 4 3 6 ENTER M550004 01...

Страница 29: ...CONSTANT CURRENT O UTPUT 10 00VO LTS 100AM PS ON OFF MENU VOLTAGE CURRENT OVERVOLT LAST SET CANCEL 1 2 3 4 Figure 3 1 DHP Series Controls and Indicators Table 3 1 2U 3U and 6U High Series Controls and...

Страница 30: ...wn the down key when the Output is displayed LIMIT SETTINGS Displayed by pressing and holding down the up key when the output screen is displayed The output settings displayed are those applied at the...

Страница 31: ...he MENU button to the Restore From Memory mode enter the desired memory location from the keypad or scroll to the location using the keys Once the desired location is selected press ENTER to retrieve...

Страница 32: ...ace For use please refer to the SCPI Option User s Manual AUTO STEP Control is passed to the auto step program Instructions on how to set up the program are in Auto step Programming Section of this ma...

Страница 33: ...ure for the power supply Fault Counts This displays recorded faults that may have occurred in the power supply This display consists of eight integers From left to right they represent Firmware This d...

Страница 34: ...tage programming return Used with J1 9 J1 15 or J1 21 and must be referenced to or within 3V of the circuit common same as pin 20 Remote on off Switch relay contacts or a direct short between this ter...

Страница 35: ...ctivate shutdown Output voltage monitor 0 10 VDC equals to 0 100 rated voltage Voltage programming return Used with J1 9 J1 15 or J1 21 and must be referenced to or within 3V of the circuit common sam...

Страница 36: ...a mating connector with remote on off jumpered for ON terminal 5 shorted to terminal 6 3 4 REMOTE CURRENT PROGRAMMING The remote current programming is used for applications that require the output c...

Страница 37: ...oltage with respect to the VP RTN J1 20 The programming current from terminal J1 21 VP RES is factory set to 1 milliamp This yields a coefficient of 1 0 of rated output voltage for each 50 ohms If mul...

Страница 38: ...closed See Figure 3 8 for connection requirements 2 Remote on off control may be accomplished by an external 12 to 120 VAC or 6 to 120 VDC or TTL CMOS source Application of AC DC or high state TTL CM...

Страница 39: ...ES COM 6 0 5 Kohms PROGRAM 5 Figure 3 3 Remote Current Programming Using Resistance 1 13 23 IP RTN COM 6 IPROG 10 0 5 VDC OR 0 10 VDC VOLTAGE SOURCE Figure 3 4 Remote Current Programming Using 0 5 or...

Страница 40: ...RES COM 6 0 5 Kohms PROGRAM Figure 3 5 Remote Voltage Programming Using Resistance 1 13 20 VP RTN COM 6 VPROG 9 0 5 VDC OR 0 10 VDC VOLTAGE SOURCE Figure 3 6 Remote Voltage Programming Using 0 5 or 0...

Страница 41: ...LOAD Figure 3 7a Remote Sensing Operation at the Load Remote Sensing Operation at the Load 80V Out 1 25 14 SENSE 12 SENSE 13 LOAD 5 6 SUPPLY OUTPUT TERMINALS J1 Figure 3 8b Remote Sensing Operation at...

Страница 42: ...1 25 14 6 ON OFF 5 Figure 3 9 Remote On Off Control by Contact Closure 25 14 ISO ON OFF 1 AC OR DC SOURCE ISO RTN 2 Figure 3 10 Remote On Off Using Isolated AC or DC Voltage Source M550004 01 3 14...

Страница 43: ...TTL CMOS 1 ISO RTN 2 Figure 3 11 Remote On Off Using Isolated TTL CMOS Voltage Supply 1 13 COM 6 REM OV SET 3 0 5 5 VDC VOLTAGE SOURCE Figure 3 12 Remote Overvoltage Set Using DC Voltage Source M55000...

Страница 44: ...This page intentionally left blank M550004 01 3 16...

Страница 45: ...s is identical Each programming sequence requires the input entries 1 The step number to be programmed 2 The output voltage for the step 3 The output current for the step 4 The OVP setting for the ste...

Страница 46: ...Figure 4 2 Auto step Programming Menu System M550004 01 4 2 SAME AS IN FIRST STEP SAME AS IN FIRST STEP UP TO NINE MAXIMUM...

Страница 47: ...ed for the current step Press ENTER AUTO STEP PROGRAMMING MODE STEP NUMBER ACTION AFTER THIS STEP DURATION OF STEP N GO TO NEXT STEP R REPEAT STEPS S OUTPUT OFF STOP CURRENT FOR THIS STEP VOLTAGE FOR...

Страница 48: ...may be stopped by pressing the CANCEL key To activate the program the control source must be change to AUTO STEP To do this press the menu key until the CONTROL SOURCE menu as seen in Figure 4 4 is re...

Страница 49: ...wer supply at full output voltage 4 Wire of sufficient size to carry the full output current of the supply 5 3 CALIBRATION PROCEDURE The calibration of the power supply is done by observing the output...

Страница 50: ...positive output of the supply and the other to the negative output of the supply Be sure to use adequately sized cable for the connections Connect the volt meter directly across the shunt at the kelvi...

Страница 51: ...ange and the calibration may proceed At this time read the output current using the volt meter and shunt and using the numeric keypad enter the correct current to five places Press ENTER when the corr...

Страница 52: ...within an acceptable range and the calibration may proceed At this time read the output voltage using the volt meter and using the numeric keypad enter the correct voltage to five places Press ENTER w...

Страница 53: ...to five places Press ENTER when the correct voltage has been input A reading of LOW instead of OK indicates either too low a resistance in the load or a faulty power supply After ENTER of the last st...

Страница 54: ...This page intentionally left blank M550004 01 5 6...

Страница 55: ...least five minutes after power has been removed via external means all circuits and or terminals to be touched must be safety grounded to the chassis 6 2 PREVENTIVE MAINTENANCE Preventive maintenance...

Страница 56: ...ed or pinched wire frayed worn or missing insulation Corrosion loose cracked or dirty connections Short circuits arcing corrosion overheating Dirt and corrosion Clean contacts with solvent moistened c...

Отзывы: