805709F / 1115
ORTEC
®
MÖBIUS™ HPGe Detector Cooling System
3
1.3.2.
Mechanical
●
Dewar Capacity
28 L, excluding detector dipstick volume.
●
Cooler
AMETEK SUNPOWER with passive vibration dampening.
●
Auxiliary Cooling
Internal fan.
●
Audible Noise
Less than 60 dB (A) at 1 m fully operational, below 30°C ambient.
●
Electrical Supply
100–240 V ac, 50 Hz or 60 Hz.
●
Normal Power Consumption
125 W typical, 350 W maximum.
●
Dimensions
(excluding HPGe detector) 25.8 in. H x 17.2 in. Diam (65.4 cm x 43.7 cm).
●
Operating Temperature
0–40°C, 20–90% relative humidity, non-condensing.
1.3.3.
CE
●
Intended for indoor use.
●
Intended for altitude up to 2000 m (6562 ft).
●
Mains supply voltage fluctuations up to +/-10% of the nominal voltage.
●
Transient overvoltages up to the levels of overvoltage category II.
●
Pollution degree 2.
1.4.
INSPECTION FOR SHIPPING DAMAGE
If the shipping carton(s) arrives with externally visible damage,
do not unpack it
. Notify the carrier and make
arrangements to file a damage claim.
In all cases of shipping damage, it is the customer's responsibility to file a
damage claim.
If, during unpacking, you find concealed damage,
notify the carrier and file a claim
. Be sure to save all packing
materials, waybills, and other such documentation to establish claims.