General use of buttons

The functions of the single buttons are only activated when you release the button. When a 
parameter is displayed, first the parameter name appears, for example 

FLO

 for flow 

AL.HI

 for 

high oxygen alarm etc. for about 1 second. Then the actual value is displayed. 

With the   button it is easy to scroll between 

FLO

AL.HI.

 and 

AL.LD

 and return to the output of 

the actual O

2

 (20.9)

No matter which parameter is currently displayed, the device automatically returns to display 
the actual O

2

 when there has been no activation of any button for about 5 seconds (the 

sensor flow can be read constantly. See 

"Indication of sensor flow" on page 13

.

Start and stop of measuring

In the set-up menu it is selected whether the measuring is controlled via the 

 button, via 

external pump control signal, button as well as external control signal or if the measuring is 
always activated. If only an external control of measuring has been chosen or the measuring is 
always activated, the button 

 is inactive. If either of the two other set-up possibilities has 

been chosen, the measuring can be started and stopped using the button 

.

1. Press 

 and the measuring starts. As soon as the measuring has started, the measured O

2

 

concentration is read out on the display.

2. Press 

 again and the measuring stops and the displays turns off (only % will be lit). 

External/manual start and stop of measuring 

If an external and manual control of measuring has been chosen in the set-up menu, the 
measuring will start when the external measuring control signal is set to ON and stop when 
the signal is set to OFF. It is only when the external control signal is changing condition that 
the measuring starts/stops. Every time the 

 key is pressed the conditions change, meaning 

that if the measuring has been started by the external control signal the measuring stops 
when 

 is activated. Now measuring can only be started by the external signal if signal is set 

to OFF and the ON again. 

IOR

$O/R

$O+L

Содержание mocon Dansensor ISM-3

Страница 1: ...P N 200114 I 09 2021 UserManual...

Страница 2: ...This blank page has been inserted to enable double sided printing of the document...

Страница 3: ...Dansensor ISM 3 EN User Manual P N 200114 I 09 2021 UserManual Published by MOCON Europe A S R nnedevej 18 4100 Ringsted Denmark Tel 45 57 66 00 88 info dk mocon ametek com www ametekmocon com...

Страница 4: ...2 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...

Страница 5: ...Set up of alarm limits 13 Indication of sensor flow 13 Flow alarm 13 Current output voltage output 13 Sub menus 14 Setting up of options relay and external measurement control 16 5 Maintenance 17 6 Ca...

Страница 6: ...4 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...

Страница 7: ...or too low O2 concentration Besides that an alarm is given in case of insufficient flow of measuring gas or in case of other faults in the ISM 3 ISM 3 is equipped with a relay for alarm in case of too...

Страница 8: ...larm limits programmable for O2 content Calibration with up to 4 calibration gases plus atmospheric air 20 946 O2 Supplied as stand alone or for incorporation ISM 3 is specifically built for measuring...

Страница 9: ...is moved or transported make sure it is note exposed to shocks or impact Only use water or a mild soap lather to clean the front panel Always make sure that there is at least 5 cm of free air behind t...

Страница 10: ...8 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...

Страница 11: ...3 Connection for measuring control signal pin 1 10 to 32VDC pin 2 GND Fuse 0 63 AT 1 6 AT at 115V AC Placed inside Fan GAS OUT GAS IN For ISM 3 without a build in pump 2 10 bar is labelled upon reques...

Страница 12: ...10 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...

Страница 13: ...me is 2 minutes The rest of the heating time in minutes is indicated in the display as H 10 H for heating When the time is counted down to zero the ISM 3 is ready After 10 minutes ISM 3 has a read out...

Страница 14: ...measuring is always activated If only an external control of measuring has been chosen or the measuring is always activated the button is inactive If either of the two other set up possibilities has b...

Страница 15: ...continuously display the flow press and hold the button for 3 sec while the flow is displayed After about 3 seconds the display will flash shortly 3 The gas flow value will be displayed until the but...

Страница 16: ...Now you have access to the 3 sub menus Set up menu SET Diagnostic menu DIA Calibration menu CAL No matter which menu you are in the device will return to display the actual O2 when no key has been act...

Страница 17: ...et the other one is displayed as If identical values are entered for the 0 and 20mA or the 4 and 20 mA values error code er 1 will be displayed Diagnostic menu DIA 1 Press when DIA appears in the disp...

Страница 18: ...rent output voltage output 1SFTT FBTVSFNFOU BDUJWBUFE POMZ WJB CVUUPO NBOVBM DPOUSPM 1SFTT YU FBTVSFNFOU DPOUSPMMFE WJB CVUUPO BOE FYUFSOBM NFBTVSJOH DPOUSPM TJHOBM YU FBTVSFNFOU BDUJWBUFE POMZ WJB FY...

Страница 19: ...analyser in a normal application situation is needed between the obligatory 12 months service control intervals See Recommendations on page 7 for details If ISM 3 is installed in places with more dust...

Страница 20: ...18 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...

Страница 21: ...ce with minimum 2 gases besides 20 9 one above and one below the range in which you want to measure Calibration NOTE ISM 3 must warm up for at least 1 hour before making the calibration The following...

Страница 22: ...larm limits on page 13 7 Press and simultaneously to start the actual the calibration The display will then flash CAL for about 10 seconds while performing the calibration Following the calibration re...

Страница 23: ...menu subsequently 50 57 Fatal hardware errors not error 54 Problem Possible cause Possible solution The display does not show O2 values even though measurement has been started ISM 3 is not warm Wait...

Страница 24: ...r two calibration gases with the same value Find the calibration gas with the same value as the one you tried to enter and calibrate The display shows er 6 An attempt has been made to calibrate with a...

Страница 25: ...tment is reset and stored in the Flash memory Calibrate the device The display shows er 33 Errors in one of or more of the set up values All set up values are changed to factory settings Please check...

Страница 26: ...SM 3 must be returned for repair The display shows er 55 Hardware error Read error from the A D Converter Switch the ISM 3 off and wait 10 seconds and then switch on again If message of error is still...

Страница 27: ...built in model 1 meter hose 5 3 mm Option voltage output ISM 3 Accessories extra Fuses 5x20mm 0 63 AT 4 pcs ser cpl P N 210548 Fuses 5x20mm 1 6 AT 4 pcs ser cpl P N 210549 Filter cooling 60x60 10 pcs...

Страница 28: ...26 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...

Страница 29: ...he flow and the measuring gas hose Calibration Atmospheric air 20 946 up to 4 optional calibration gases Flow system For ISM 3 without an internal pump the measuring gas is supplied at a pressure of 2...

Страница 30: ...dard For measuring in other ranges device must be recalibrated see Calibration on page 19 for details Possible connection to the ISM 3 Gas supply Gas inlet on the back Relay Max 48V 1A COMMON N O and...

Страница 31: ...C Less than 95 RH up to 35 C during operation Linear RH reduction by 3 RH C between 35 C and 40 C Permissible ambient temp 20 C to 45 C Less than 75 RH between 35 C and 45 C while stored IP classifica...

Страница 32: ...30 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...

Страница 33: ...This blank page has been inserted to enable double sided printing of the document...

Страница 34: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...

Отзывы: