System 4
Target Orbiting Thermometer
User Supplement
© Land Instruments International, 2008-2014
Issue 5
28 March 2014
Publication Nº 198.069
Language: English
Страница 1: ...System 4 Target Orbiting Thermometer User Supplement Land Instruments International 2008 2014 Issue 5 28 March 2014 Publication N 198 069 Language English...
Страница 2: ...n It is possible that this equipment is to be installed on or near to machinery or equipment operating at high temperatures and high pressures Suitable protective clothing along with face and eye prot...
Страница 3: ...ystem of Land Instruments International is approved to BS EN ISO 9001 for the design manufacture and on site servicing of combustion environmental monitoring and non contact temperature measuring inst...
Страница 4: ...Target Orbiting Thermometer User Supplement...
Страница 5: ...arget Orbiting Thermometer Contents 1 About this Supplement 1 2 System 4 Target Orbiting Thermometers 1 2 1 Supplementary Specification 1 3 Principle of Operation 3 3 1 Suggested settings for use in P...
Страница 6: ...4 M2 Thermometer 0 25 K 1K Accuracy for a Target Orbiter System 4 M1 Thermometer 0 5 K 1K Temperature C Specification C M2 M2 TO M1 M1 TO 300 0 Not Specified Not Specified 400 0 2 7 4 4 500 0 2 9 4 9...
Страница 7: ...eld Of View M1 600 1600 CV TO 100 1 M1 600 1600 CS TO 100 1 M1 1100 2900 FV TO 100 1 M1 1100 2900 FS TO 100 1 M1 800 2600 CV TO 200 1 M1 800 2600 CS TO 200 1 M1 1500 4700 FV TO 200 1 M1 1500 4700 FS T...
Страница 8: ...arget spot diameter D 100 D 200 for high temperature versions rad D 29 D Distance to Target Plane d target spot diameter identical to a standard thermometer 2 Optical Field of View D d Fig 1 Field of...
Страница 9: ...intermittent targets In a standard System 4 Thermometer the eyepiece graticule defines the target spot In a Target Orbiting Thermometer the graticule defines the orbit of the centre of the target spo...
Страница 10: ...erature above which the Peak Picker is active The lowest selectable level is the span minimum of the thermometer connected but otherwise is decided entirely by the user 3 Set the On delay to zero 4 Se...
Страница 11: ...truments based on System 4 UNO thermometers accessories and mountings and special instruments based on UNO Application dedicated processors based on LANDMARK Graphic ABTS S and ABTS U FTS VDT S and VD...
Страница 12: ...ns outside the Ambient Temperature Humidity and Vibration rating of the product environmental contamination solvent vapours deposition of airborne contamination cooling liquids of non neutral pH etc o...