3.4.3 Setting output voltage range
Select 100 V or 200 V. The output voltage range cannot be changed while output is on.
The setting range varies according to the output voltage range.
Table 3-5. List of Setting Ranges for Various Output Voltage Range
Setting
Setting Range
100 V range
200 V range
Unit
DC
voltage
200.0 to +200.0
400.0 to +400.0
V
AC
voltage
Wave
-form
SIN/SQU
0.0 to 140.0
0.0 to 280.0
Vrms
ARB1 to 16
0.0 to 400.0
0.0 to 800.0
Vp-p
Output peak current limiter
(positive)
+10.0 to +42.0
+5.0 to +21.0
A
Output peak current limiter
(negative)
42.0 to
10.0
21.0 to
5.0
A
Output average current limiter
1.0 to 10.5
1.0 to 5.3
A
Positive voltage setting limit
+5.0 to +200.0
+10.0 to +400.0
V
Negative voltage setting limit
200.0 to
5.0
400.0 to
10.0
V
External input gain
0.0 to 200.0
0.0 to 400.0
times
Operation steps
To set the output voltage range, select
in the SET menu.
1.
Press the
or
key to move the cursor to the
icon.
ENTER
CANCEL
Cursor
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...
Страница 248: ......
Страница 266: ......
Страница 274: ......