4.3 Measurement of Harmonic Current
Switching power sources are widely used in both consumer and industrial electrical products. The
capacitor input type rectifier circuits that are used with such input power lines have disadvantages such
as significant distortion of input current and a large amount of harmonic current. When a large amount
of this current flows in the power source line, the equipment may experience operation faults, such as
voltage distortion, which can cause the transformer to overheat, possibly resulting in a hazardous
accident.
The EC1000S includes a function for measuring harmonic current. This measurement does not comply
with the IEC standards.
This measurement can be performed only in AC-INT mode and when the fundamental wave (panel
frequency setting) is either 50 Hz or 60 Hz. Absolute values of harmonic current in RMS and their
percentages to the fundamental are displayed up to the 40th harmonics (2 kHz at 50 Hz fundamental).
When harmonic current measurement is selected, RMS value, average DC, and peak value
measurements are not performed. If RMS value, average DC, or peak value has been set, harmonic
current measurement is not performed.
a) Changing to harmonic measurement window
To view the harmonic measurement window,
select “MEASURE” in the MISC menu
.
Operation steps
1.
Press the
or
key to move the cursor to the
icon, then press the
E NT E R
key.
ENTER
CANCEL
Cursor
The MISC menu is displayed.
2.
Press the
or
key to move the c
ursor to “MEASURE”, then press the
E NT E R
key.
The selection box is displayed.
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...
Страница 248: ......
Страница 266: ......
Страница 274: ......