PN 9000-165-VE, Rev P
10. The screens will advance upon completion of each calibration to the next
gas to be calibrated.
11. When all gases have calibrated, the screen displays the Calibration Results
that include
% drift
for gases calibrated as well as recovery time. If the
calibration is successful, a large GREEN check mark will be on the screen. If
it is not, a large RED ‘X’ is displayed.
Perform a Remote Calibration
The functions for remote calibration will only be visible if you have and
RCU and you have configured your sensor for
Automatic Calibration
in the
Sensor Configuration
menu.
Using the Remote Cal Unit (RCU), you can manually accept the gas values
when each gas is stable by pressing the
ENTER/MENU
key, but the RCU
switches the gases on/off automatically.
1. From the
Main Menu
screen, press
ENTER/MENU
to access the
Calibra-
tion Menu
.
2. Press the Left or Right arrows to locate the
Start Calibration
menu and
press
ENTER/MENU
.
3. Navigate to the drop down box, press
ENTER/MENU
key, and choose
Remote
.
4. Press the Right arrow to continue.
5. Select the gas(es) that you want to calibrate.
6. Enter the time for the Recovery Duration in minutes.
7. Press the Right arrow to continue.
This is the
Calibration Gas Values
screen. If you have already entered
your calibration gas values in the
Cal Gas Menu
, they will be visible in
the fields. Any changes that you make to these values will be reflected
in the
Cal Gas Menu
as well.
8. When you are satisfied with your values, press
ENTER/MENU
to start the
remote calibration. The screen displays the
Gas Value
,
Cal Gas
, and
Status
(
Ready
/
Wait
). To abort the calibration, select the
ABORT
button and press
ENTER/MENU
.
2-24 | Thermox
®
AMEVision Display User Interface
Содержание AMEVision
Страница 1: ...AMEVision Display User Interface USER MANUAL PN 9000 165 VE Rev P 9000 185 VE Manual Views Rev A ...
Страница 12: ...PN 9000 165 VE Rev P This page intentionally left blank xii Thermox AMEVision Display User Interface ...
Страница 46: ...PN 9000 165 VE Rev P This page intentionally left blank 2 28 Thermox AMEVision Display User Interface ...
Страница 70: ...PN 9000 165 VE Rev P This page intentionally left blank 4 4 Thermox AMEVision Display User Interface ...
Страница 78: ...PN 9000 165 VE Rev P WDG V AMEVision Modbus Map A 2 Thermox AMEVision Display User Interface ...
Страница 79: ...PN 9000 165 VE Rev P EventMask 1 Bit definitions AMEVision Modbus Information A 3 ...
Страница 80: ...PN 9000 165 VE Rev P WDG 5 Command Definitions A 4 Thermox AMEVision Display User Interface ...
Страница 81: ...PN 9000 165 VE Rev P Status Mask Bit definitions AMEVision Modbus Information A 5 ...
Страница 82: ...PN 9000 165 VE Rev P Process Alarm Bit Definitions A 6 Thermox AMEVision Display User Interface ...
Страница 83: ...PN 9000 165 VE Rev P EventMask 2 Bit definitions AMEVision Modbus Information A 7 ...
Страница 84: ...PN 9000 165 VE Rev P This page intentionally left blank A 8 Thermox AMEVision Display User Interface ...