![Ametek 933S Скачать руководство пользователя страница 247](http://html1.mh-extra.com/html/ametek/933s/933s_operators-manual_3044056247.webp)
Maintenance and Troubleshooting
|
6-71
Microcontroller Board Alarm Conditions and Correc-
tive Action
This section lists valid alarms (errors) that originate from the
Microcon-
troller
board, the Bit and Decimal Value for each Alarm Type (Fault or
Warning), descriptions of the alarms, and suggested corrective actions.
The letter ‘
f
’ preceding the alarm condition indicates it is a Fault, while ‘
w
’
indicates it is a Warning.
Microcontroller alarms are displayed automatically by the Configurator
Software under
MC Scode
on the
Status
tab (Figure 6-15). If no alarm con-
dition exists, nothing is displayed.
Decimal Alarm
Bit
Value
Type
Alarm Condition / Description and Suggested Corrective Action
0
1
Warning
w: PMT signal out of range
The signal from either the
Measure
or
Reference
PMT is outside its normal range
(2500 mV–9840 mV). View these signals from the
Optical Bench Data
tab.
Corrective Action:
Take appropriate safety precautions, open the Upper Enclosure, and:
•
Check the flat cables between the
Optical Bench
board and Optical Bench
for proper connections and inspect them for damage (cuts, nicks, burn marks,
etc.).
•
Initiate an Auto-Setup.
See “The Auto-Setup Sequence” in this chapter.
•
Replace the lamp(s), if necessary (see “Source Lamp Replacement” in this
chapter to determine if the lamps need to be replaced and how to replace
them).
•
Replace the
Optical Bench
and/or
PMT Buffer
board.
Contact AMETEK to
verify operation before removing these boards.
Содержание 933S
Страница 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 34: ...2 10 Model 933S UV Analyzer...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 933S UV Analyzer...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Страница 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Страница 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Страница 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Страница 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Страница 187: ...Maintenance and Troubleshooting 6 11 Figure 6 2 Optical Bench Assembly with Standard Range Measuring Cell...
Страница 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Страница 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Страница 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Страница 229: ...Maintenance and Troubleshooting 6 53 Figure 6 14 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Страница 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 268: ...A 4 Model 933S UV Analyzer Sample Flow Diagram Plumbing Schematic North American Style WX 933NA 2...
Страница 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Страница 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Страница 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Страница 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Страница 285: ...Appendix A Drawings A 21 Inlet and Vent Flame Arrestor Locations Environmental Purge Option 100 1783 1A...
Страница 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...