Maintenance and Troubleshooting
|
6-49
Chopper Assembly Maintenance
Replace the Chopper Motor Drive Belt and Bearings in the Chopper As-
sembly every 2 years.
When the Ex d Upper/Lower Enclosure is opened for maintenance or
repair, take extreme care to avoid scratching or damaging its flame-
paths.
If at any time the enclosure door is open and service personnel are
not working on or in the Upper/Lower Enclosure, close the door and
secure it with at least one screw. This will reduce the risk of inadver-
tently scratching or damaging the flamepath.
Before closing the door, gently clean the flamepath with a soft, non-
abrasive cloth and make sure the surfaces are free of debris.
“P/N” refers to Part Number.
To replace the Drive Belt and Bearings in the Chopper Assembly (see Fig-
ures 6-13 and 6-14):
Hazardous Locations
Before proceeding, test the area around the analyzer for explosive
gases and proceed only when the area is found to be safe.
Do not remove the cover of the Explosion-Proof Digital Communica-
tions Port on the front of the Lower Enclosure, do not connect a serial
cable to the Communications Port, do not open the Upper/Lower En-
closures, and do not power up/down the analyzer or computer if there
is an explosive gas atmosphere present.
1. Connect a serial cable between the computer and the analyzer’s Com-
munications Port. Using the Configurator Software, change the
Flow
Control
setting to ‘
3
’ (
zero
). Click
OK
then
Apply
. Do not save the
changes to EEPROM.
!
CAUTION
NOTE
!
WARNING
See Chapter 7 for
spare part ordering numbers.
Setup (tab)
Gas Calibration
<<Flow
Control>>
Содержание 933S
Страница 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 34: ...2 10 Model 933S UV Analyzer...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 933S UV Analyzer...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Страница 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Страница 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Страница 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Страница 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Страница 187: ...Maintenance and Troubleshooting 6 11 Figure 6 2 Optical Bench Assembly with Standard Range Measuring Cell...
Страница 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Страница 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Страница 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Страница 229: ...Maintenance and Troubleshooting 6 53 Figure 6 14 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Страница 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 268: ...A 4 Model 933S UV Analyzer Sample Flow Diagram Plumbing Schematic North American Style WX 933NA 2...
Страница 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Страница 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Страница 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Страница 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Страница 285: ...Appendix A Drawings A 21 Inlet and Vent Flame Arrestor Locations Environmental Purge Option 100 1783 1A...
Страница 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...