5-4 | Model 931S / Model 932S UV Analyzers
Manual Span Gas Calibration
A Manual Span allows you to Span calibrate individual gas species. The
analyzer span factors are adjusted automatically at the completion of the
Span. See also “Setting Gas Calibration Parameters” in Chapter 4.
To perform a Span gas calibration:
1. From the
Gas Calibration
dialog box, click
AutoZero
to perform an
Auto-Zero to automatically adjust the analyzer zero. The zero offset is
adjusted based on the average reading during the
Integration Timer
duration at the end of the
Auto Calibration Timers
(
example, AZ-0
)
countdown.
2. After the Auto-Zero process is complete, record the Zero gas pressure
setting and then turn off the Zero gas.
3. Remove the Swagelok cap from the Manual Calibration Gas Injection
fitting and connect the Span gas tubing. See Figure 5-3 for a general
layout and refer to the Overall Component Layout drawing in the
Documentation Package shipped with the analyzer.
4. Adjust the Span gas pressure to the recommended pressure listed on
the system drawings. The pressure must be less than the 1380 kPag
(200 psig) maximum Cell Pressure.
5. Change the
Flow Control
setting to ‘
3
’ (
span1
) or ‘
4
’ (
span2
) then
click
Apply
(next to
Flow Control
text field). Click
OK
to return to the
Setup
tab but do not save the changes to EEPROM.
6. Change the span concentration (
Span Conc
) value of the gas spe-
cies to be calibrated to the actual Span gas value that is being used.
Click
OK
to return to the
Setup
tab but do not save the changes to
EEPROM.
7. From the
Analyzer Data
tab, observe the Span Concentration read-
ings. After the readings have stabilized, return to the
Gas Calibration
dialog box and click the
ManSpan
button for the gas species being
calibrated. To check the time remaining for the span function, observe
the
Time to end of cal(sec)
timer under
Countdown timers
on the
Analyzer Data
tab. You can also verify that the span calibration is in
progress by viewing
Calibration/Solenoid Status
field in the upper-
right corner of the
Status
tab. It should display “
Manual-span1
” (or
“
Manual-span2
” for
Model 932S Analyzers only
).
At the completion of the Span, confirm that the Span Factor (
SFactor
,
Gas Calibration
dialog box) for the species has been updated. The
default value for Span Factor is 1.0.
Setup (tab)
Gas
Calibration
Содержание 931S
Страница 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 34: ...2 10 Model 931S Model 932S UV Analyzers...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 931S Model 932S UV Analyzers...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 931S Model 932S UV Analyzers...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 931S Model 932S UV Analyzers...
Страница 71: ...Installation and Start Up 3 25 Figure 3 8 Conduit entries and AC power connection locations Lower Enclosure...
Страница 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 221: ...Maintenance and Troubleshooting 6 37 Figure 6 12 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Страница 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Страница 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Страница 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Страница 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Страница 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Страница 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...