Maintenance
— 42
SPECTROLAB — 23.07.2020
7.5
Cleaning the exhaust filter system and changing the filter
insert
Warning! Toxic or flammable substances!
The sparking process produces a black precipitate (metal condensate). The
condensate could be toxic or flammable, depending on the sample material.
Protective gloves and safety goggles must always be worn when cleaning the
spark stand and when cleaning the gas-washing bottle or the filter bottle. This
also applies when replacing the filter insert. If metals with toxic components
have been sparked off, wearing a breathing mask to prevent the inhalation of
condensate particles is mandatory.
When disposing of the condensate and the used filters, compliance with legal
regulations is mandatory.
Fig. 25
The exhaust gas filter system must be checked and cleaned at regular intervals, if
necessary—at the latest when the software reports a high pressure of the
operating gas (see image). If the pressure of the operating gas is too high, the
condensate produced during the measurement is not discharged sufficiently and
accumulates in the spark stand. It may also be necessary to clean the spark stand
as described in the chapter 7.4.
1. Open the door by pulling the
side lever (1) and swinging the
door to the right.
Fig. 26