Cleaning and Replacing Parts
”PN” in the following procedure(s) refers to AMETEK Part Number.
Systems With an AMETEK-Supplied Sample Panel
The Filter (membrane separator), which is located on the AMETEK Sample
Panel, protects the analytical system (Sample Cell, Flow Meter, etc.) from con-
tamination by dirt, dust, and liquids that can be present in the process stream.
Because process stream contaminants can vary from stream to stream there is
no single replacement interval that is right for all applications and installations.
AMETEK recommends that you inspect the Filter Membrane frequently (e.g.,
weekly) for new installations, to determine how long it typically takes for the
Filter to become dirty. If the Filter remains clean for extended periods of time,
the Filter can be checked less frequently, if preferred.
Figure 4-1.
5100P Analyzer with AMETEK-
supplied Sample Conditioning
Panel (Standard 250 PSIG
version shown).
FI
SAMPLE PANEL
TOP VIEW
INLET VALVE
REGULATOR
GENIE FILTER
DRAIN VALVE
FLOWMETER
WITH VALVE
OUTLET FITTING
(TO INSTRUMENT)
GAUGE
MAX PANEL INLET PRESSURE: 250 PSI
INSTRUMENT INLET/PANEL OUTLET: 101 KPA +/- 2.5 KPA
FLOW RATE: 1-2 SLPM
REGULATOR
INLET VALVE
GAUGE
DRAIN
VALVE
GENIE
FILTER
FLOWMETER
WITH VALVE
DRAIN
SAMPLE PANEL
FLOW SCHEMATIC
5100P
ANALYZER
To check and replace the Filter Membrane:
1. Disconnect all process gas and vent connections from the analyzer and
completely depressurize the Filter Assembly.
2. Turn Off the analyzer power.
3. Remove the knurled Filter Cap from the Filter Assembly by hand, and set it
aside. The cap should not require a wrench to unscrew it.
The Filter Membrane is mounted to the inside of the cap with the o-ring.
4-8 | 5100P TDLAS Analyzer
Содержание 5100P
Страница 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Страница 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Страница 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Страница 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Страница 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...