Hoefer DALT System
Hoefer DALT System Function and Description
Amersham Biosciences
11
Required or Convenient, But not Supplied
Refrigerated Cooling Bath
DALT Tank cooling is provided by circulating chilled liquid through the heat exchanger
on the floor of the central chamber. With the aid of Quick-fit connectors, you can
connect a refrigerated circulating bath, such as the MultiTemp III, to the fittings at the
back of the DALT Tank. We strongly recommend active cooling for protein transfers.
Focused IPG Strips
Immobiline DrySrips provide a stable gradient pH gradient (IPG) immobilized onto an
acrylamide matrix and supported by a plastic film backing. They are readily available,
easy to handle and not prone to breakage and stretching. IPG strip focusing is
accomplished on the IPGphor or Multiphor II with the Immobiline DryStrip tray
accessory.
Power Supply
For overnight or full-day runs, a conventional electrophoresis power supply delivering
600 V at 400 mA, such as the EPS 601, is suitable. For multiple tanks or faster runs (8 –
9 h) a higher current power supply, such as the EPS 2A200, is desirable.
Transfer Membranes
The DALT Blotting kit includes 50 pieces of blotting paper. For electrophoresis tank
blotting, Amersham Biosciences offers a complete line of transfer membranes,
including pure nitrocellulose, supported nitrocellulose, PVDF and nylon. Choose the
appropriate transfer membrane for your application.
Small Gadgets
•
A small plastic spatula or thin plastic ruler
to insert and seat IPG gels on the tops of
the slab gels between the glass plates
•
Plastic-covered dish racks (Rubbermaid or equivalent)
for holding the DALT gel
cassettes during the loading process
•
A microwave oven or 100 °C heating block
for preparing the agarose sealing
solution
•
A Wonder Wedge (80-6127-88)
for prying open the gel cassettes after the DALT run
•
Screw-cap culture tubes, 25
×
200 mm, for storing IPG strips at –40 °C
•
GelSeal for lubricating the gradient gasket, to assure a leakproof seal
•
A peristaltic pump
is recommended for more rapid removal of buffer than the small
siphon pump supplied with the DALT Tank. The pump can be set up to deliver used
buffer that is not radioactive directly to a lab drain.
•
A vacuum pump
for degassing solutions
•
Thick, latex dishwashing gloves, with textured fingers