TROUBLESHOOTING GUIDE
LAB STORAGE TANKS
25
98-0123 Rev E
10
TROUBLESHOOTING GUIDE
Symptom
Possible Cause
Correction
High Bacteria
Counts
UV light quartz sleeve
dirty
Clean sleeve with scale cleaner
Replace quartz sleeve
Water Appears
“Milky”
Air in water line
Run water until air is purged
“Lamp Life
Expired” Error on
UV Controller
Lamp expired
Replace UV lamp
“Sensor Failure”
Error on UV
Controller
No signal to UV controller
Ensure sensor cable is properly connected to
controller
Disconnect sensor for 5 seconds to re-set sensor
Replace UV sensor
“Interlock Open”
Error on UV
Controller
Lamp connector not fully
seated
Ensure that the lamp connector is fully seated
Verify locking clip is in place
“Lamp Failure”
Error on UV
Controller
Lamp is not providing
signal to the controller
Verify engagement of the lamp to the lam
connector
Lamp connection contacts are corroded
Lamp expired - replace
“Low Level UV”
Error on UV
Controller
Insufficient UV for
adequate disinfection
Clean quartz sleeve and sensor lens
Replace UV Lamp
Replace UV sensor
Red Alarm Light
Illuminated on
Pump
Pump has an alarm
condition
Pump has ran dry – add water to storage tank
Pump is over temperature – allow to cool then re-
start
Overloaded motor – will attempt to restart every
30 minutes for a 24-hour period. Replace if not
corrected.
Seized up pump - will attempt to restart every 30
minutes for a 24-hour period. Replace if not
corrected
Содержание 00HC-4090
Страница 2: ......
Страница 7: ...COMPONENT IDENTIFICATION LAB STORAGE TANKS 5 98 0123 Rev E 2 COMPONENT IDENTIFICATION...
Страница 8: ...COMPONENT IDENTIFICATION LAB STORAGE TANKS 6 98 0123 Rev E...
Страница 9: ...COMPONENT IDENTIFICATION LAB STORAGE TANKS 7 98 0123 Rev E 2 1 Flow Schematic...
Страница 15: ...OPERATION LAB STORAGE TANKS 13 98 0123 Rev E Figure 3 Grundfos Eye Indicators...
Страница 16: ...OPERATION LAB STORAGE TANKS 14 98 0123 Rev E 5 2 Default Pump Program...
Страница 28: ...REPLACEMENT PARTS LAB STORAGE TANKS 26 98 0123 Rev E 11 REPLACEMENT PARTS...