7
098-0007 Rev B
2.4
Residual Dangers
WARNING:
The following dangers may be present if operating the system with the side covers removed.
Electrical Shock: With the side covers removed, the control system for the device will be
exposed. 120 VAC, 24 VDC and 12 VDC power supplies are all contained within the device. Take
necessary precautions to avoid shock when servicing the or maintaining the system.
Mechanical Force: With the side covers removed and the system operating, some parts of the
system may be under pressure up to 70 PSI.
Hot Surfaces: During heat disinfection, some of the internal surfaces will become hot to touch.
Take extra care if performing a disinfection cycle with the side covers removed.
2.5
Handling
This device is intended to be used as a stationary device. Do not move the device while in operation.
Prior to moving the device, refer to the decommissioning procedure listed in Section 4.11,
Decommissioning for relocation. The dry weight of the system is 57 lbs. Take care when lifting the
system on or off of the transport cart.
When moving the system, be sure the water tank is empty. To manually drain the water tank if full, refer
to Section 7.5.2.
2.6
Bringing the unit to an Immediate STOP
If you need to stop the unit immediately at any time, switch the rocker switch located at the rear of the
system off at any time. As a precaution, shut the RO system down to shut off the water supply to the
Centurion Plus Heat disinfection device. Refer to the RO manual for detailed information on bring the RO
to an immediate stop.
DO NOT restart the unit until it has been verified to be safe for operation. If the original fault that
required the unit to be shut down cannot be resolved, contact AmeriWater for assistance. Contact
information can be found in Section 1.3.
2.7
Dealing with leaks from the unit
In the event that a leak forms inside the unit, shut the system down using the rocker switch on the rear
of the unit. If the leak occurs while the unit is in “Standby” mode, the RO can be disconnected from the
Centurion Plus Heat disinfection device to continue patient treatment until a certified technician can
address the leaking component.
If water leaks outside of the system, the water should be cleaned immediately. If the water cannot be
cleaned immediately, it is strongly recommended that a “Caution Wet Floor” sign is placed near to device
to warn people of a potential slip hazard.
2.8
Unauthorized conversion and manufacturing replacement parts
DO NOT under any circumstances modify, or replace parts inside of the device with unauthorized parts
or attempt to change or alter the functionality of the device. Installation of unapproved components may
void the warranty of the device
WARNING:
The Centurion Plus Heat disinfection system should be used in accordance with Section 1.2,
Intended Use. The device is to be maintained and operated according to this operation
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...