background image

 

August 18, 2003 

INSTALLATION, OPERATION AND MAINTENANCE 

INSTRUCTIONS FOR THE AMI

 

MODEL 150A

 

LIQUID HELIUM LEVEL METER

 

I.

 

INTRODUCTION 

The American Magnetics, Inc. (AMI) Model 150A is a compact, fully portable Liquid Helium 
Level Meter providing simple and easy operation.  The instrument is battery operated and when 
used in conjunction with an AMI LHe Level Sensor, provides a truly portable means of 
measuring liquid helium levels at any location.  The sensor connects to the instrument via a 
standard DB9 connector.  The instrument displays 0-100% of sensor maximum active length 
immersed in LHe.  The instrument can be used with AMI sensors with active lengths of up to 60 
inches.  The instrument is powered by sealed lead-acid batteries.  A standard AC-DC adapter 
(either 120VAC or 230VAC) is supplied with the unit for charging the batteries.  Since battery 
energy consumption is strongly dependent on the LHe level being measured, AMI has developed 
a special "battery saver" circuit which reduces the power drain from the batteries as more of the 
sensor becomes superconducting. Power requirements decrease as LHe level increases.  A front 
panel POWER LED is provided to indicate sufficient battery power exists to make an accurate 
level reading. 

The CE mark applies to the handheld unit and the European AC adapter.  The North American 
AC adapter is not CE marked. 

II.

 

SPECIFICATIONS (at 25 °C unless stated otherwise) 

Input Power (instrument) 

Sealed lead-acid battery (6VDC)

 

Input Power (AC adapter)

 

90VAC to 132VAC 50/60 Hz, or 180VAC to 264VAC 
50/60 Hz  (depending on specified AC adapter model) 

Operating temperature (instrument)

 

0 to 50 °C non-condensing

 

Operating temperature (AC adapter)

 

0 to 40 °C non-condensing

 

Sensor active length range 

1 inch to 60 inches 

Linearity

 

0.5% of full scale

 

Absolute Calibration

 

0.5% of full scale 
2% of full scale (from 0 to 50 ºC)

 

Probe Current

 

75 mA nominal

 

Battery Life

 

Typically 1,000 readings with a 60" sensor.

 

* Under extreme radiated electromagnetic field conditions (3V/m @ 80MHz to 1GHz) the 
accuracy may be degraded by an addi/- 1.2% 

Содержание 150

Страница 1: ...AMI has developed a special battery saver circuit which reduces the power drain from the batteries as more of the sensor becomes superconducting Power requirements decrease as LHe level increases A front panel POWER LED is provided to indicate sufficient battery power exists to make an accurate level reading The CE mark applies to the handheld unit and the European AC adapter The North American AC...

Страница 2: ...ilitate mobility the Model 150A is equipped with a single standard DB9 connector which connects the instrument to a sensor Connect the LHe sensor wires to the DB9 connector terminals as follows Wire Color DB9 Pin Number Signal Name Red 1 I Yellow 6 V Black 7 I Blue 8 V IV OPERATION A Connect the instrument to the sensor via the sensor DB9 connector Refer to Figure 1 B Depress the READ push button ...

Страница 3: ...3 August 18 2003 Fig 1 Model 150A ...

Страница 4: ...sed NOTE The AC adapter is rated for connection to Category II mains voltages NOTE Only AMI approved AC adapters should be used on the device NOTE Only an authorized service representative should replace the factory installed batteries C As part of the routine maintenance keep the exterior surfaces of the instrument clean by gently wiping with a damp cloth moistened with a mild detergent VI CALIBR...

Страница 5: ...ibration resistor Jumper the I and V contacts pins 6 and 7 on the sensor connector together and connect to the other side of the calibration resistor The value of the calibration resistor should be 4 56 ohms x the active length of the sensor in cm remember that the active length of AMI standard sensors is 1 inch less than the overall length 2 Temporarily install a short circuit using a clip lead o...

Страница 6: ...nnected in accordance with the color code outlined in the installation section see table in installation section and Figure 1 If the cause of the problem cannot be located please call an AMI Technical Support representative at 865 482 1056 Our standard business hours are Monday through Friday 8 00 a m to 4 30 p m Eastern Time IV QUALIFIED SERVICE PERSONNEL QSP A ESD This instrument contains compon...

Страница 7: ...the printed circuit board then disconnect them from the printed circuit board 5 Remove the READ switch cap and guard cap is pulled off and guard is unscrewed 6 Apply slight pressure to the POWER LED and READ switch from the front of the Model 150A and gently remove the printed circuit board from the back NOTE Gently removing the printed circuit board will ensure that the POWER LED disengages from ...

Страница 8: ...perform as specified for a period of one year from date of shipment In the event of failure occurring during normal use AMI at its option will repair or replace all products or components that fail under warranty and such repair or replacement shall constitute a fulfillment of all AMI liabilities with respect to its products Since however AMI does not have control over the installation conditions ...

Страница 9: ...l 150 Standards to which Conformity is Declared Safety EN 61010 1 2001 EMC EN55 022 2003 Class A EN61326 2002 EN61000 4 2 2002 8kV AD 4kV CD EN61326 2002 EN61000 4 3 2002 3V m 80 1000MHz EN61326 2002 EN61000 4 8 2001 30A m 50Hz I the undersigned hereby declare that the equipment specified above complies with the require ments of the aforementioned Directives and Standards and carries the CE mark a...

Отзывы: