36
Rev. 5
Operation
Single-key Commands : Ramp / Pause
three seconds before reverting to the default display. The value displayed
is as follows:
•When in driven mode, the present current/field will be displayed.
•When in persistent mode, the current/field will be displayed that
was flowing in the magnet at the time persistent switch was
cooled.
3.6.3
Ramp / Pause Key
Pressing the
RAMP / PAUSE
key toggles the Model 430 Programmer
between the ramping mode and the paused mode. If the
RAMP / PAUSE
key is pressed while the Model 430 is ramping, the ramping is paused. If
the
RAMP / PAUSE
key is pressed while the Model 430 is paused, the
Model 430 continues ramping.
3.6.4
Ramp To Zero Key
Pressing the
RAMP TO ZERO
key causes the Model 430 Programmer to
immediately begin ramping field/current up or down to zero field/current
at the defined ramp rate(s). Ramping to zero may be interrupted at any
time by pressing the
RAMP / PAUSE
key, which causes the Model 430
Programmer to enter the PAUSED mode and maintain the field/current
present at the point it was paused.
Note
If the
RAMP TO ZERO
function is PAUSED and then the
RAMP /
PAUSE
button is pressed a second time, the Model 430 Programmer
will begin ramping to the target field, not to zero. If it is desired to
ramp to zero after the
RAMP / PAUSE
button is pressed, press the
RAMP TO ZERO
button again to continue ramping to zero from the
paused state.
Note
If the magnet is persistent (persistent switch heater is not energized)
when the
RAMP TO ZERO
key is pressed, the Model 430
Programmer ramps the power supply current to zero; the magnet
current will remain constant since the magnet is in persistent mode.
+0.25 A
P
Magnet Current (A)
+0.00 Vs
+10.00 A
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...