112
Rev. 5
Remote Interface Reference
System-Related Commands
4.5 Command Reference
The following paragraphs present all Model 430 Programmer commands
and queries in related groups and a detailed description of the function of
each command or query is provided. Examples are also provided where
appropriate. Return strings may be up to 80 characters in length.
4.5.1
System-Related Commands
•
*IDN?
Return the identification string of the Model 430 Programmer. The
identification string contains the AMI model number and firmware
revision code.
•
*RST
Resets the Model 430 Programmer. This is equivalent to cycling the power
to the Model 430 Programmer using the power switch. All non-volatile
calibration data and battery-backed memory is restored. Status is cleared
according to the
*PSC
setting.
•
*TST?
Performs a self-test. Currently always returns “1”.
•
<
Ctrl-C
>
This clears the output buffers of the Model 430 Programmer and prepares
the instrument for a new command. Status registers are unaffected.
<
Ctrl-
C
>
corresponds to ASCII code 03.
•
SYSTem:LOCal
Enables all front panel controls. All front panels controls are enabled by
default after a power-up or
*RST
command.
•
SYSTem:REMote
Note
The SYSTem:REMote command only disables the front panel
controls for purposes of preventing accidental operation of a front
panel feature. It is
not
necessary for this command to be sent prior
to using a remote interface. Send the SYSTem:LOCal command,
send the *RST command, press
SHIFT
followed by
MENU
, or cycle
Model 430 Programmer power to re-enable the front panel controls.
Disables all front panel controls. If the Model 430 Programmer is in the
remote mode, an asterisk (
*
) will appear in the front panel display in the
position just below the ramping character as shown below.
Содержание 05100PS-430-601
Страница 2: ......
Страница 10: ...x Rev 5 List of Figures ...
Страница 12: ...xii Rev 5 List of Tables ...
Страница 18: ...xviii Rev 5 Foreword Safety Summary ...
Страница 30: ...12 Rev 5 Introduction Operating Characteristics ...
Страница 42: ...24 Rev 3 Installation Power Up Procedure ...
Страница 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Страница 119: ...Rev 5 101 Remote Interface Reference SCPI Command Summary LOCK ABsorber LOCK BRIGHTness LOCK NETsetup ...
Страница 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Страница 168: ...150 Rev 5 Service Return Authorization ...
Страница 190: ...172 Rev 5 Appendix Power Supply Details Figure A 3 Model 08150PS Dimensions Top and Side Views ...
Страница 220: ...202 Rev 5 Appendix Persistent Switch Operation Flowchart Figure A 17 Persistent Switch Operation Flowchart Page 3 ...
Страница 226: ...208 Rev 5 Index ...