36
-
Error Number
Problem
Remedy
E L
(Lid lock trouble
alarm)
・
The sliders for locking the lid are not
inserted at the right position.
・
The detection of locking turn off during
operation.
・
Contact our authorized distributor in
your region.
2. Early Troubleshooting
Symptom
Cause
Remedy
Display remains off
after power is
turned on.
●
Check the plug and outlet first.
①
The plug is not properly inserted or
is insufficiently tightened.
②
Disconnection in the power cord.
③
Defect in the DISPLAY.
④
The main power switch is "OFF."
①
Properly insert the plug and
retighten any loose parts.
②③
Contact our authorized distributor
in your region.
④
Turn "ON" the main power switch.
No air exhausted
from the working
chamber.
①
Defective automatic exhaust valve.
①
Contact our authorized
distributor in your region.
Pressure gauge
reading remains
low.
①
Defective safety valve.
②
Defective pressure gauge.
③
Disconnection in the heater.
④
Defective automatic exhaust valve.
⑤
Steam leakage.
①
-
④
Replace the defective part
(Contact our authorized distributor in
your region).
⑤
For steam leakage from piping,
retighten or seal the joints.
Steam leakage
from lid gasket
①
Deterioration of lid gasket
②
Improperly installed lid gasket.
③
Foreign matter under the gasket.
①
Replace the lid gasket.
②
Press on the gasket to remove any
unevenness.
③
Remove the foreign matter.
Water leakage from
the bottom of the
body.
①
Deterioration of the heater seal
packing due to boil-dry or other
problem.
②
The drain valve open.
①
Contact our authorized distributor in
your region.
②
Close the valve.
The lid will not
open.
①
The lid open display is red.
②
Power is not supplied.
①
Wait until the lid open display
changes to green.
②
Press the "POWER" switch to turn
on the power.
The lid will not
close.
①
The lid is not open completely.
①
Press the "LID OPEN" switch to
completely open the lid, and then
press the "LID CLOSE" switch.
Displayed
temperature
exceeds set
temperature and
exhaust is repeated
frequently during
the sterilization
stage.
①
Defect in the heater circuit.
①
Contact our authorized distributor in
your region.
●
This table of early troubleshooting describes the causes and remedies of simple problems.
If you are unable to fix or repair the problem, contact our authorized distributor in your region and
provide the following information. (Please show the warranty to our service staff visiting your
place.)
①
Model and serial number of the autoclave.
②
Defective point(s) and symptom(s) (error number if applicable).
③
Number of days of operation (date of purchase).
④
Operating conditions (including substances being sterilized).