4. Selection of a Mode and Program
●
Select a mode according to the usage.
①
Press the mode switch (LIQ, AGAR, SOLID, or
DISSOL).
・
Every time the switch is pressed, the current mode
changes from Mode 1 to Mode 2, 3, 1,
・・・
in sequence repeatedly.
Mode
Application
Liquid
Sterilization of water, media, solid reagent, liquid reagent, liquid chemical, and
other liquid that can withstand high-temperature high-pressure steam.
Agar
Sterilization of agar media (to be warmed so that it will not solidify after sterilization)
Solid
Sterilization of glass, ceramics, metal, rubber, and other solid equipment that can
withstand high-temperature high-pressure steam and sudden decompression
during exhaust
Dissolution
Dissolution of media
Liquid mode
Stage display
Initial set value (for all 3 programs)
Sterilizati
on temp.
Sterilizati
on time
Exhaust
level
Fan
cooling
Warm
temp.
HEAT
→
STERI
→
EXHT./COOL
121
℃
20 min.
0
OFF
Agar mode
Stage display
Initial set value (for all 3 programs)
Sterilizati
on temp.
Sterilizati
on time
Exhaust
level
Fan
cooling
Warm
temp.
HEAT → STERI →
EXHT./ COOL → WARM
121
℃
20 min.
0
OFF
50
℃
AGAR
LIQ.
SOLID
DISS
OL