8 |
ENGLISH
CLEANING
Cleaning the Ameda Platinum Breast Pump
For Personal Use
After each pumping, turn OFF the pump. Unplug cord from from wall outlet. Wipe the Ameda Platinum Breast
Pump with clean, damp cloth.
For Hospitals/Rental Stations
Turn OFF the pump. Unplug cord from wall outlet. Clean between users. Wearing disposable gloves, wipe the
Ameda Platinum Breast Pump with a hospital approved antibacterial cleaner, such as CaviCide
®
, CaviWipes
®
,
or PDI
®
Sani-Cloth
®
AF3 .
STORING AND TRANSPORTING
Disconnect HygieniKit and power cord from pump before transporting or storing the product.
When storing product, store in conditions as described in these instructions.
Handle with care when transporting or handling product.
TROUBLESHOOTING
Pump does not turn ON
• Ensure proper power cord connection to pump.
• Ensure grounded plug connection to electrical outlet.
• Ensure outlet has power .
Unable to adjust speed
• Ensure pump function is ON.
• Ensure “Check Kit” is not shown in suction display window.
Unable to adjust suction
• Ensure pump function is ON.
• Ensure “Check Kit” is not shown in suction display window.
Low or no suction
• Ensure “Check Kit” is not shown in suction display window.
• Check breast flange for adequate seal against breast.
• Check that the Ameda HygieniKit Milk Collection System is properly assembled.
• Check that the adapter cap is firmly snapped (attached) onto top of breast flange.
• Ensure silicone diaphragm is in place inside top of breast flange.
• Ensure tubing adapter is properly fitted into tubing adapter port on pump.
• Inspect valve for stretching or tears. If seen, replace valve.
• Ensure valve is pushed firmly into place in lower portion of breast flange.
• Ensure one side of tubing adapter is closed when single pumping .
“Check Kit” error message appears
• Ensure tubing adapter is properly fitted into tubing adapter port on pump.
• Ensure one side of tubing adapter is closed when single pumping .
• Check that adapter cap is firmly snapped onto top of breast flange.
• Ensure silicone diaphragm is in place inside top of breast flange.
• Check tubing and silicone diaphragm to ensure that parts are free of defects .
• Check tubing is firmly attached to adapter cap.
“Service Required” error message appears
• Call your local distributor. For listing of distributors worldwide, visit www.ameda.com.
The Ameda Platinum Breast Pump should be serviced only by Ameda, Inc. or authorized Service Centers
approved by Ameda, Inc.