![Amann Girrbach 181364 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/amann-girrbach/181364/181364_instruction-manual_2918061018.webp)
18
S P A R E A N D W E A R P A R T S
E N
▷
Mount the transparent splash guard to the
extraction opening and lock it in place with the
lever.
2.5
Changing a calibration sleeve
The blank holder for rotational milling is designed
such that the calibration sleeve can be replaced
and exchanged.
Fig. 8
▷
To remove the calibration sleeve [3] from the
tool holder [1], loosen and completely unscrew
both screws [4] using the size 3 Allen key.
▷
Pull off the calibration sleeve [3].
▷
Insert a new calibration sleeve with the
uncoated side onto the tool holder.
▷
Screw the calibration sleeve firmly to the tool
holder.
3
Spare and Wear Parts
Exchanging the calibration sleeve is only
necessary in case of an error in the cali-
bration process. When exchanging the cal-
ibration sleeve, it is required to carry out
the calibration process once more.
1
3
2
4
Both the calibration sleeve [3] as well as
the tool holder [1] have a total of four
holes.
▷
Screw the calibration sleeve to the tool
holder only via the horizontal holes [2].
The vertical holes of the tool holder are
used to affix the blank holder (see
chapter 2.4.5 on page 17).
The screws of the calibration sleeve are
shorter than those for the blank holder.
▷
Do not switch the screws or get them
mixed up.
Art. No.
Spare part
179279
Calibration sleeve
179288
Blank screw and blank nut
760672
Roto RFID 2,0 Ti
760673
Roto RFID 1,0 Ti
Tab. 2