Example:
If
Log3
is selected at the
REC
function the analog output will be spread over 3
decades of measurement. If
0.1
is chosen as the low starting point at the
REC Lo
function then
the first decade of measurement will be from
0.1
to
1.0
, the second from
1.0
to
10.0
and the
third from
10.0
to
100.0
. For a 4-20mA output the output for each decade would be as shown
in the table below:
Display range
mA output range
0.1 to 1.0
4.0 to 9.3
1.0 to 10.0
9.3 to 14.6
10.0 to 100.0
14.6 to 20.0
5.55
Second analog output logarithmic or linear mode
Display:
REC2
Range:
Lin, Log1, Log2, Log3, Log4
or
Log5
Default Value:
Lin
Second analog output linear or logarithmic mode - seen only when the dual analog output option
is fitted and not applicable unless
REC ctrL
is set to
OFF
. See
REC
for a description.
5.56
Second analog output logarithmic start point
Display:
REC2 Lo
Range:
0.001, 0.01, 0.1, 1.0
or
10.0
Default Value:
0.001
Second analog retransmission logarithmic output low starting point - Seen only when the dual
analog retransmission option is fitted and not applicable unless
REC ctrL
is set to
OFF
. See
REC Lo
for a description.
5.57
Baud rate for optional serial communications
Display:
bAUd RAtE
Range:
300, 600, 1200, 2400, 4800, 9600, 19.2
or
38.4
Default Value:
9600
Set baud rate - seen only with serial output option. Refer to the separate “PM4 Panel Meter
Optional Output Addendum” booklet supplied when optional outputs are fitted.
Select from
300, 600, 1200, 2400, 4800, 9600, 19.2
or
38.4
baud. The baud rate should be set to
match the device being communicated with.
5.58
Parity for optional serial communications
Display:
PrtY
Range:
NONE, EUEN
or
odd
Default Value:
NONE
Set parity - seen only with serial output option. Refer to the separate “PM4 Panel Meter Optional
Output Addendum” booklet supplied when optional outputs are fitted. Select parity check to
38 of