that the valley memory value is being displayed.

HiLo

- toggle between

Hi

and

Lo

displays. This function allows the remote input to be used

to toggle between peak and valley memory displays. The first operation of the P button will
cause the peak memory value to be displayed, the next operation will give a valley memory
display.

PHi

or

PLo

will flash before each display to give an indication of display type.

5.21

Access mode

Display:

ACCS

Range:

OFF, EASY, NONE

or

ALL

Default Value:

OFF

Access mode - the access mode function

ACCS

has four possible settings namely

OFF, EASY,

NONE

and

ALL

. If set to

OFF

the mode function has no effect on alarm relay operation. If set

to

EASY

the “easy alarm access” mode will be activated. Refer to “Easy alarm relay adjustment

access facility” section. If set to

NONE

there will be no access to any functions via

FUNC

mode,

entry via

CAL

mode must be made to gain access to alarm and calibration functions. If set to

ALL

then access to all functions, including calibration functions, can be gained via

FUNC

mode.

5.22

Number of input bits

Display:

INPt bitS

Range:

1

to

20

or

1

to

31

Default Value:

14

Input bits - allows selection of the number of input bits from 1 to 20 for parallel inputs or 1 to 31
for strobed or addressed inputs.

5.23

Sign bit on or off

Display:

SIGN

Range:

on

or

OFF

Default Value:

OFF

Sign bit

on

or

OFF

(parallel binary input only) - Example: if set to on then, with 16 bit binary

selected, the most significant digit will be used as the sign bit with 0 on the most significant digit
being positive and 1 on the most significant digit being negative. e.g. 0111 1111 1111 1001 will
show as

32761

on the display and 1111 1111 1111 1001 will show as

-7

(-7 being the twos

compliment of 111 1111 1111 1001). Thus the display with 15 bits plus a sign bit can display any
number from

-32768

to

32767

. If set to

OFF

none of the data bits will be treated as a sign

bit and all values will be positive.

PM4BCMAN-2.0-0

29 of

40

Содержание PM4-BC

Страница 1: ...splay Controller Operation and Instruction Manual AMALGAMATED INSTRUMENT CO ABN 80 619 963 692 Unit 5 28 Leighton Place Hornsby Telephone 61 2 9476 2244 e mail sales aicpl com au NSW 2077 Australia Fa...

Страница 2: ...cal Installation 4 3 Electrical installation 5 4 Function tables summary of setup functions 16 5 Explanation of functions 19 6 Scaled inputs 35 7 Weighted input operation 37 8 Specifications 39 9 Guar...

Страница 3: ...high ambient light levels The high contrast LCD displays provide good visibility and are ideal for battery powered applications Full electrical isolation between power supply input signals and retrans...

Страница 4: ...n option for panel mount instruments Prepare a panel cut out of 45mm x 92mm 1 mm 0 mm see diagram below Insert the instrument into the cut out from the front of the panel From the rear of the instrume...

Страница 5: ...terminal plug in data connector is provided which will accept up to 1mm2 wires To insert wires to this data connector insert a screwdriver blade max 2 5mm wide into the square socket adjacent to the...

Страница 6: ...Instrument label example 3 3 Resistor pack and link locations 3 4 Electrical connection examples 6 of 40 PM4BCMAN 2 0 0...

Страница 7: ...3 5 Parallel BCD input PM4BCMAN 2 0 0 7 of 40...

Страница 8: ...3 6 Strobed BCD input 8 of 40 PM4BCMAN 2 0 0...

Страница 9: ...3 7 Addressed BCD input PM4BCMAN 2 0 0 9 of 40...

Страница 10: ...3 8 Parallel binary input also used for Gray code 10 of 40 PM4BCMAN 2 0 0...

Страница 11: ...3 9 Strobed binary input also used for Gray code PM4BCMAN 2 0 0 11 of 40...

Страница 12: ...3 10 Addressed binary input also used for Gray code 12 of 40 PM4BCMAN 2 0 0...

Страница 13: ...set at the dCPt function will be ignored The table below shows the input requirements for remote decimal point selection The display hold input will hold the current display value when activated The I...

Страница 14: ...3 12 Weighted digital input See chapter 7 for description of this input type 14 of 40 PM4BCMAN 2 0 0...

Страница 15: ...pping screws from the back cover then remove the back cover by pulling it away from the instrument 3 Remove the earth screw which passes through the underside of the case then slide out the board or b...

Страница 16: ...y x Optional 0 to 9999 0 See 4 1 5 4 23 Axrt Reset time delay for the designated alarm relay x Optional 0 to 9999 0 See 4 1 5 5 23 Axn o or Axn c Alarm relay x action to normally open de energised or...

Страница 17: ...al Any display value 1000 5 18 28 dCPt Decimal point 0 0 1 etc 0 5 19 28 Pbut P button function for instruments with front P button NONE Hi Lo or HiLo NONE 5 20 28 ACCS Access mode OFF EASY NONE or AL...

Страница 18: ...e for serial communications Optional 300 600 1200 2400 4800 9600 19 2 or 38 4 9600 5 40 33 PrtY Parity for serial communications Optional NONE EUEN or odd NONE 5 41 33 O Put Output for serial communic...

Страница 19: ...emove power from the instrument Hold in the button and reapply power The display will briefly indicate as part of the wake up messages when the message is seen you can release the button Move to step...

Страница 20: ...e on to the next setpoint Note this easy access also functions in the same manner for the PI control setpoint relay and or analog PI output if PI control is available The instrument must be set in the...

Страница 21: ...w and high setpoints Example If A2Lo is set to 10 then relay 2 will activate when the display value is 10 or less Display Value Time A Lo x A HY x value A Lo A HY x x plus Relay activates at this valu...

Страница 22: ...around setpoint value e g if A2HY is set to zero the alarm relay 2 will activate when the display value reaches the alarm setpoint for high alarm and will reset when the display value falls below the...

Страница 23: ...nds binary Gray code whole number BCD bcd COdE 0 and weighted digital inputs only note there is no relay 1 therefore relay numbering starts from 2 The reset time is common for both alarm high and low...

Страница 24: ...a positive number and if operating behind the prime setpoint then the value is entered as a negative number Example With Alarm relay 3 set to trail alarm 2 if A2Hi is set to 10 0 and A3Hi is set to 5...

Страница 25: ...t supplied when this option is fitted for wiring details and link settings Displays and sets the analog retransmission 4 20mA 0 1V or 0 10V link selectable output low value 4mA or 0V in displayed engi...

Страница 26: ...uld also be set up as a centre zero double dot display by entering a negative value and positive value e g bAr_ 100 bAr 100 C bAr centre bar display The display will be a solid bargraph but will have...

Страница 27: ...ption BCD start position Display bcd Strt Range 0 1 or 2 Default Value 0 Seen only with the 16 bit digital output option Refer to the separate PM4 Panel Meter Optional Output Addendum booklet supplied...

Страница 28: ...ed to be recalibrated and alarm etc settings checked For addressed strobed and 4 digit parallel BCD inputs the decimal points can be selected remotely via the signal levels on the rear input terminals...

Страница 29: ...made to gain access to alarm and calibration functions If set to ALL then access to all functions including calibration functions can be gained via FUNC mode 5 22 Number of input bits Display INPt bi...

Страница 30: ...fault Value 0 Weighting for digital input 1 2 etc seen only when SEt INPt is set to d inP Allows weighted values to be set for each of the digital inputs used May be set to any value in the display ra...

Страница 31: ...1010 to 1111 displays blanks 1 A to F 1010 to 1111 displays A b C d E F 2 A to F 1010 to 1111 displays C 1 o blank 3 A to F 1010 to 1111 displays A L H I o blank 4 A to F 1010 to 1111 displays C F _...

Страница 32: ...terminal 21 cannot be used to hold the display See HOLd POL function below 5 35 Remote display hold polarity Display HOLd POL Range Hi or Lo Default Value Lo Polarity for display hold input for addres...

Страница 33: ...t Value 9600 Set baud rate seen only with serial output option Refer to the separate PM4 Panel Meter Optional Output Addendum booklet supplied when optional outputs are fitted Select from 300 600 1200...

Страница 34: ...al serial communications Display Addr Range 0 to 31 Default Value 0 Set unit address for polled POLL or M buS mode 0 to 31 seen only with serial output option Refer to the separate PM4 Panel Meter Opt...

Страница 35: ...e plus 10 e g 160 Press the F button to accept the change The display scaling will operate in parallel strobed or addressed modes scaling cannot be used with BCD inputs The electrical inputs for scale...

Страница 36: ...he input to 111001100100 3000 decimal then go to the CAL1 function and press and v simultaneously A display value will be seen 2 Press the F button the message SCL1 will be seen followed by a value Us...

Страница 37: ...g retransmission The display is to have two decimal points and each weighted input in turn is to be given a value of 1 00 3 00 5 00 and 10 00 mA The input signals are to be from a switch inputs with a...

Страница 38: ...nactive Active Inactive 5 00 9mA Active Inactive Active Inactive 6 00 10mA Inactive Active Active Inactive 8 00 12mA Active Active Active Inactive 9 00 13mA Inactive Inactive Inactive Active 10 00 14m...

Страница 39: ...it 12 7mm 6 digit 12 7mm Power Supply AC 240V 110V or 24V 50 60Hz or DC isolated wide range 12 to 48V Special supply types 32VAC 48VAC 50 60Hz or DC isolated 50 to 110V also available Note supply type...

Страница 40: ...representative of the manufacturing company Products for attention under guarantee unless otherwise agreed must be returned to the manu facturer freight paid and if accepted for free repair will be re...

Отзывы: