5.18
Digital output option high value
Display:
diG~
Range:
Any display value
Default Value:
1000
Seen only with the 16 bit digital output option. Refer to the separate “PM4 Panel Meter Optional
Output Addendum” booklet supplied when this option is fitted. Determines the high scaling point
for the
b.SCL
mode and has no effect on other modes.
Example:
If
diG_
is set to
0
and
diG~
is set to
65535
(2
16
−
1) then the retransmission
will not be scaled i.e. a display of
2
will cause a retransmission of 2. If
diG~
is now changed to
32767
(2
15
−
1) then a display of
2
will cause a retransmission of 4 (note: rounding may occur
on retransmission).
5.19
Decimal point
Display:
dCPt
Range:
0
,
0.1
etc.
Default Value:
0
Displays and sets the decimal point. By pressing the
^
or
v
pushbutton at the
dCPt
function
the decimal point position may be set. The display will indicate as follows:
0
(no decimal point),
0.1
(1 decimal place),
0.02
(2 decimal places),
0.003
(3 decimal places) and
0.0004
for display
with more than 4 digits. Note if the decimal point is altered the display will need to be recalibrated
and alarm etc. settings checked.
For addressed, strobed and 4 digit parallel BCD inputs the
decimal points can be selected remotely via the signal levels on the rear input terminals 18, 19 and
20. See section
and
dP inPt
function for details. When remote decimal point selection is
used the
dCPt
function settings are ignored.
5.20
P
button function
Display:
Pbut
Range:
NONE, Hi, Lo
or
HiLo
Default Value:
NONE
P
button function - a only applicable models with front panel
P
buttons. The
P
button is located
at the front of 5, 6 or 8 digit LED models and bargraph models.
Functions available are:
•
NONE
- no function required.
•
Hi
- peak memory. The peak value stored in memory will be displayed when the
P
button
is pressed, if the button is pressed momentary then the display will return to normal mea-
surement after 20 seconds. If the button is held for 1 to 2 seconds or the power is removed
from the instrument then the memory will be reset.
PHi
will flash before the reading to
indicate that the peak memory value is being displayed.
•
Lo
- valley memory. The minimum value stored in memory will be displayed. Otherwise
operates in the same manner as the
Hi
function.
PLo
will flash before the reading to indicate
28 of
PM4BCMAN-2.0-0
Содержание PM4-BC
Страница 7: ...3 5 Parallel BCD input PM4BCMAN 2 0 0 7 of 40...
Страница 8: ...3 6 Strobed BCD input 8 of 40 PM4BCMAN 2 0 0...
Страница 9: ...3 7 Addressed BCD input PM4BCMAN 2 0 0 9 of 40...
Страница 10: ...3 8 Parallel binary input also used for Gray code 10 of 40 PM4BCMAN 2 0 0...
Страница 11: ...3 9 Strobed binary input also used for Gray code PM4BCMAN 2 0 0 11 of 40...
Страница 12: ...3 10 Addressed binary input also used for Gray code 12 of 40 PM4BCMAN 2 0 0...
Страница 14: ...3 12 Weighted digital input See chapter 7 for description of this input type 14 of 40 PM4BCMAN 2 0 0...