4 | Perform Measurements
BYK-Gardner GmbH
40
spectro2guide Pro
295 024 855 E 2106
Continue with
Next
and place instrument on the
white
surface.
Click on
Measure
or press the
Operate
button.
Sample is measured and automatically saved.
Scroll down to see the
Opacity
value in %.
4.5 Fluorescence Measurement
The
spectro2guide
and the
spectro2guide Pro
can detect fluorescent light ex-
cited in UV and visible range. With this function you can predict the color stability.
A monochromatic illumination is used to excite fluorescent material in the sample.
For this the instrument is equipped with additional 12 LEDs (360 - 660 nm):
• 2 ultra-violet LEDs and
• 10 monochromatic LEDs.
To measure fluorescence, the corresponding indices have to be activated. Go to
Configuration
>
Measurement parameter
>
Fluorescence
.
Following indices are relevant: