1.0 General Description & Features
1.1 Description
The
dewTEC
DS2000 is a DIN style panel mounted dewpoint meter, designed to accurately
measure the dewpoint in gas or dry compressed air. The operating range of the unit is factory
set to the Moisture Sensor supplied with the unit. Process moisture value in factory set units
of °C, °F or PPM(V) is indicated, by the
dewTEC
DS2000 dewpoint meter, on a large five
digit seven segment display
The main external body of the Ultra High Capacitance Moisture Sensor is made of plated steel
and a stainless steel filter guard protects the sensing element within it.
The sensor can be located up to one kilometre away from the dew point meter.
1.2 Main
Features
1.2.1 Dewpoint
Meter
Features
The
dewTEC
DS2000 dewpoint meter’s main features are:
•
Display: Five large LED digits and one small LED digit, both with seven segments.
Two alarm status LED’s.
•
User programmable LED indication to display in either Green or Red and can be
programmed to change colour when in alarm condition.
•
Programmable via push button keys on front panel or PC (optional extra).
•
Alarms: Two, independent, fully user programmable isolated relays with open
collector outputs. Programmable lock out to avoid unauthorised tamper.
•
Displays maximum and minimum values.
•
Displays elapsed time from alarm condition.
•
Factory set calibration data in °C, °F or PPM.
•
Factory set current (mA) OR voltage (mV) analogue output. Isolated and user
programmable to span over any range, within the overall range of the sensor.
•
RS485 serial output with 1-99 address settings.
•
90 - 264V, 50/60Hz Universal input power supply.
•
Accuracy
±
2°C dewpoint (or equivalent in other moisture units).
•
AUTOCAL feature for automatic calibration adjustment.
1.2.2 Sensor
Features
The sensor main features are:
•
Various dewpoint ranges available. Range ident on sensor packaging.
•
Conditioned for long term stability.
•
Can be located up to 1km from the
dewTEC
DS2000 dewpoint meter, connected by a
low impedance coaxial cable.
•
ATEX Certified for use in potentially explosive areas with suitable Zener Barrier.
dewTEC
DS2000 User’s Manual