4.2.  General settings (P1)

Menu P1 allows to adjust the general functions and settings of the 
force gauge. 

The menu offers the following submenus:

P11      Select dimensional units

P12 

Ch/- prefix for push or pull operations

P13  

Select display-update frequency

P14 

Automatic measuring stop 

P15  

Automatic switch off

P16 

Automatic taring

Select the submenus one after the other by pressing the O-key. 

With the I-key you select the settings in the submenus. The actual set  
value flashes on the display will be set by leaving the submenu at 
this point. You can leave the submenu by selecting the next submenu 
pressing the O-key or by reverting to the menu on the level above.

By pressing the S-key you revert to the level above. 

By pressing  the S-key for 2 seconds you revert to the operation level.

4.2.1.  Selecting dimensional units (P11)

The SI-unit Newton (N) is set by default.

Submenu P11 allows to change the measuring unit displayed. Press  
the I-key to navigate between the options. The selected unit flashes 
on the display.

 

[Measuring range up to 10 N:  

N

 - cN - g - oz ]

[Measuring range up to 50 N:  

N

 - kg - lb ]

N

N

 cN 

(Instruments with measuring range up to 10 N)

 kg 

g

(Instruments with measuring range up to 10 N)

lb

oz

(Instruments with measuring range up to 10 N)

Example setting parameters: 
P11 - Dimensional units

The selected measuring unit flashes on the display

> <

> <

> <

> <

> <

> <

P11   

Units

N

P12   

Prefix

 

P13   

Display-refresh rate

P14   

Auto-Stop

P15   

Auto-Off

P16   

Auto-Tara

Selecting submenus in the menu P1 General settings

Содержание FMI-B Series

Страница 1: ...ucciones de servcio Istruzioni per l uso Alluris GmbH Co KG Subject to changes and corrections without notice 2016 04 13_EN Operation Manual Force Gauges FMI S FMI B FMI B FMI S Deutsche Version Engli...

Страница 2: ...S P14 17 4 2 5 Automatic measuring stop Auto Stop for FMI B P14 18 4 2 6 Auto Off function instrument off P15 18 4 2 7 Automatic taring function Auto Tara P16 18 4 3 Resetting factory defaults PO 19 4...

Страница 3: ...ay be can destroy the internal sensor Please consider the technical data in regard to the environmental conditions see p 30 chapter 7 0 Technical Data The instrument is equipped with a temperature com...

Страница 4: ...s de 2 1 Scope of supply Standard scope of delivery part no o Optional accessories not included FMI B10 FMI S10 FMI B20 FMI S20 FMI B30 FMI S30 FMI B50 FMI S50 Protection case Base unit Type label wit...

Страница 5: ...the instrument for more than 2 month in the dark you should expose it to light for at least 8 hours in the regu lar working environment 250 Lux before use Or you may use the USB cable to charge the i...

Страница 6: ...SO Motor control Drive unitcontrol FMT 220 MK USB 2 0 Mini B socket Function Accessory part Part number Data transfer to MS Excel1 Software FMI_Connect FMI 972 Data analysis on PC Software FMI_Analyze...

Страница 7: ...after long storage in the dark 6 10 pole Hirose socket series FMI B30 und FMI S30 and higher for digital I O s service calibration etc 7 I key On Off press 2 seconds Start Stop measurement select next...

Страница 8: ...not use tools to screw adapters onto the instrument s threaded axle Radial and side forces can da mage the instrument As gauges are often used in the scope of destruction tests or for appraising secur...

Страница 9: ...on instrument off P15 Force gauges with LiPo battery series FMI B measure upon each start the remaining capacity If it is less than about 4 hours bat 2 is dis played on the screen In this case charge...

Страница 10: ...pressing the S key 3 3 1 Standard mode In standard mode the actual measured value is displayed By default the display refresh rate is 10 Hz In the settings menue P13 you may change the refresh rate s...

Страница 11: ...ent indicates the peak value for tension operations You may change the prefix see p 17 chapter 4 2 2 Changing prefix for push or pull operations P12 Instruments series FMI B20 and FMI S20 and higher c...

Страница 12: ...top function see p 17 chapter 4 2 4 Automatic measu ring stop Auto Stop for FMI S P14 allows to stop a measurement automatically if measured values remain stable within a defined period of time The fe...

Страница 13: ...shows the mea sured real time value In all of the three peak modes the instrument captures and stores beside the actual measured value also the posi tive and negative peak value In Peak mode or Peak...

Страница 14: ...d values can be deleted one after the other by pressing the O key The display shows O O after having deleted a value By pressing the S key you can select the next stored value and proceed in the same...

Страница 15: ...c settings menus P1 General settings see p 16 chapter 4 2 General settings P1 P2 Memory FMI B30 and FMI S30 and higher see p 19 chapter 4 4 Memory and statistic functions P2 P3 Upper limit FMI B30 and...

Страница 16: ...enu by selecting the next submenu pressing the O key or by reverting to the menu on the level above By pressing the S key you revert to the level above By pressing the S key for 2 seconds you revert t...

Страница 17: ...ency of 3 6 kHz To ensure the legibility of the display the display refresh rate is limited by default to 10 Hz Submenu P13 allows to reduce this value further Press the I key to select the desired fr...

Страница 18: ...evel 4 2 6 Auto Off function instrument off P15 Force gauges running without USB cable are turned off if no button is pressed for more than five minutes as long as you are not working in the settings...

Страница 19: ...S key Reset factory defaults 0 no Reset 1 Reset 4 4 Memory and statistic functions P2 4 4 1 General considerations on memory functions Force gauges series FMI B30 and FMI S30 and higher are able to c...

Страница 20: ...es Series Pressing the S key for 2 seconds the instrument returns to the opera tion level in idle status To confirm activated memory the display now shows the according Mem symbol in addition to the s...

Страница 21: ...individual values You may pause the recording of measurement values while the measuring keeps running and restart the recording both by pressing the S key Recording ends with the termination of the me...

Страница 22: ...djacent chart Standard mode is set last measured value S S S S S S S S S S N N Mem Peak modes are set Peak last measured value Mem Peak N N N Peak positive peak value N N Mem Peak negative peak value...

Страница 23: ...hen by software Devices of series FMI S have to be connected to PC already during measurements 4 4 7 Clearing memory You have three possibilities to clear the memory Single peak values can be deleted...

Страница 24: ...As soon as the measured value is smaller than the lower limit value a downward pointing arrow is shown on the right side of the display Accordingly an upward pointing arrow is shown if the measured v...

Страница 25: ...oad or the exceeding of limit values via USB 2 0 or Hirose cable to a PC or a motor driven test stand 5 1 Data transfer via USB For data tranfer via USB you need a USB cable part no FMI 931USB and a s...

Страница 26: ...FMI 976 offers basically all analyzing functions as FMI_Analyze Furthermore the software packa ge allows to control the optional motor for test stands series FMT 220 and enables the setting of parame...

Страница 27: ...or 100 mm FDM 250C1 with a distance of 25 mm respectively 45 mm to the reference mark 5 2 Hirose Socket Force gauges series FMI B30 and FMI S30 and higher have a 10 pole Hirose socket for several ext...

Страница 28: ...GND 3 Gr n Green DIGIN1 Digital input 4 Gelb Yellow TRIGGER_OUT C Trigger signal measuring frequency 5 Grau Grey TRIGGER_OUT E 6 Orange Orange DIGOUT1 C Digital output setting via FMI_Analyze 7 Blau B...

Страница 29: ...sed materials operating forces of control elements or adhesive forces of packaging ma terials or foils exacly and reproducible Please do not hesitate to contact us for more comprehensive appli cation...

Страница 30: ...0 15 Temperature offset Tk relative K 0 02 O Point offset To absolute K 0 02 Permissible error of measurement 0 5 0 3 Tracking To surpressed Auto Tara automatic compensation Overload output USB interf...

Страница 31: ...xcluded from this warranty Besides the warranties as agreed upon in our general terms and conditions shall apply 8 2 Product registration In order to receive information about product changes and upda...

Страница 32: ...he dark for a longer period Yes Even after a self discharging of the internal energy storage the instrument can be used immediately A back up battery guarantees the basic function of the device see p...

Страница 33: ...protection laws would have to be considered as free A 2 Calibration confirmation acc DIN EN 10204 2 1 We hereby certify that the equipment has been tested in the production process in accordance with...

Страница 34: ...79100 Freiburg Germany Type Number FMI S10xx FMI S20xx FMI S30xx FMI S50xx FMI B10xx FMI B20xx FMI B30xx FMI B50xx see type label Description Digital Force Gauge Serial number See type label Applicab...

Отзывы: