Advanced parameters and commands
Goldeye Technical Manual
V2.2.0
77
Baud rate (s)
The baud rate of the RS232 interface is determined by use of register
s
:
The bits of the parameter value are divided into three different sized bit groups
a
,
e
and
x
. Each bit group has a special function, being independent of the other
ones:
•
a
: Controls the configuration of the serial port.
•
e
: Controls the echo mode.
•
x
: Controls the baud rate of the serial port.
Caution
Do not change the baud rate, if you control the camera via GigE
Vision features.
s 2A
Baud rate
F
E
D
C
B
A
9
8
7
6
5
4
3
2
1
0
-
-
-
-
-
-
-
-
e
a
a
-
x
x
x
x
rw
rw rw
rw rw rw rw
Bit group Value (hex)
Description
x
O
110 Baud
1
300 Baud
2
600 Baud
3
1200 Baud
4
2400 Baud
5
4800 Baud
6
9600 Baud
7
19200 Baud
8
38400 Baud
9
57600 Baud
A
115200 Baud
a
00
RS232 via 15 pin SUB-D only
01
Frame grabber‘s serial port (Camera Link or Pleora device port serial 0)
10
Additionally activate second channel via Pleora device port bulk 0
(MODE:UART) GigE interface only.
11
Reserved
e
0
Each character received from the host is echoed back to it.
1
No Echo
Table 42: Baud rates: values