Basic parameters and commands
Goldeye Technical Manual
V2.2.0
62
Temperature warning (T)
The command
T=1
displays the content of the temperature warning register. The
lowest order bit has the same function as LED L3 placed at the backside of the
camera housing. In case the value 1 is output, the temperature of the InGaAs
sensor is outside the optimum range.
When you switch on the cold camera, wait some minutes. The sensor tempera-
ture needs this time to stabilize to the set value.
In case that the temperature warning remains active for a longer period, check
the ambient temperature of the camera.
Software version and correction data
information (V)
The command
V=1
shows information of the correction data installed in the
non-volatile memory. The current firmware version and also the serial number
of the camera are output.
GigE feature name (CameraSpecialFeatures)
Feature visibility Description
SensorTemperatureState
Expert
Camera sensor temperature
state.
0 = The sensor temperature is
OK.
1 = The sensor temperature is
outside the optimum range.
SensorTemperatureStateReg
Expert
Camera sensor temperature
state register.
QuerySensorTemperatureState
Beginner
Query camera sensor
temperature state.
(T=1 command)
Table 29: GigE feature: Camera sensor temperature
Note
This function is not available as GigE Vision feature.