Allied Healthcare VACUTRON CONTINUOUS-INTERMITTENT Скачать руководство пользователя страница 1

   

VACUTRON

 SUCTION REGULATORS 

Vacuum Regulators for Continuous or Intermittent Drainage 

OPERATION AND MAINTENANCE MANUAL 

 

 

 

 

 

CONTINUOUS-INTERMITTENT 

MODEL 

INTERMITTENT 

MODEL 

PEDIATRIC CONTINUOUS-INTERMITTENT 

MODEL 

 

 
 

THIS OPERATION AND MAINTENANCE MANUAL APPLIES TO THE 

CONTINUOUS-INTERMITTENT, INTERMITTENT, AND PEDIATRIC 

CONTINUOUS-INTERMITTENT MODELS

 

 
 
 
 

CAUTION 

FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY 

OR ON THE ORDER OF A PHYSICIAN. 

 
 

WARNING 

READ THE ENTIRE MANUAL PRIOR TO USE, MAINTENANCE OR REPAIR

 

 

 

 

 
 
 
 
 

S168-319-001 

Rev. P 

 

 

 

Continuous/Intermitt

ent 

OFF 

  INT 

    REG 

OFF 

  INT 

 

Intermittent 

OFF 

 INT 

    REG 

Pediatric 

Содержание VACUTRON CONTINUOUS-INTERMITTENT

Страница 1: ...ITTENT MODEL THIS OPERATION AND MAINTENANCE MANUAL APPLIES TO THE CONTINUOUS INTERMITTENT INTERMITTENT AND PEDIATRIC CONTINUOUS INTERMITTENT MODELS CAUTION FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY...

Страница 2: ...2 SYMBOL DEFINITION READ DECREASE COUNTER CLOCKWISE INCREASE CLOCKWISE...

Страница 3: ...termittent ON and OFF Time 5 Overall Dimensions 5 Weight 5 3 0 Assembly and Set Up 6 4 0 Operating Procedure 7 4 1 Regulated Vacuum 7 4 2 Intermittent Vacuum 7 5 0 Troubleshooting Guide 7 6 0 Maintena...

Страница 4: ...el in the REG continuous and INT intermittent modes Vacuum limiting relief valves for the continuous and intermittent modes Independent variable ON and OFF vacuum cycles in the INTERMITTENT mode Quiet...

Страница 5: ...0 to 300 mm Hg Increments of 5 mm Hg 0 to 20 kPa Increments of 2 kPa 0 to 40 kPa Increments of 2 kPa Size 2 Inch Diameter Gauge Face 2 Inch Diameter Gauge Face Type Accuracy 3 of full scale Diaphragm...

Страница 6: ...suction canister connects to the bottom of the Vacutron regulator either directly or through a hose Some typical suction systems are shown in Figures 1 and 2 Figures 1 2 CAUTION DO NOT DIRECTLY ATTACH...

Страница 7: ...r low suction a Low vacuum supply b Aspirant in unit c Loose or faulty connection d Worn or obstructed regulator seat e Clogged filters f Stuck relief valve Check hospital vacuum level Flush unit Set...

Страница 8: ...lief valves These valves have the relief ball at the top of the valve If you suspect a malfunction of this style relief valve replace it Later models have adjustable relief valves These relief valves...

Страница 9: ...ontaining isopropyl alcohol 6 6 2 Turn the mode selector switch to the desired mode and turn the vacuum level knob lower large knob clockwise as far as it will go for maximum flow Draw approximately 1...

Страница 10: ...g 6 8 4 Turn the mode selector switch to the INT position 6 8 5 The timing valves are located partially under the gauge See Figure 9 Early models have white thumb wheels for valve adjustment Later mod...

Страница 11: ...11 Clockwise Increases The Time And Counterclockwise Decreases The Time ON Timing Control OFF Timing Control Timing Module Figure 9...

Страница 12: ...12 7 0 VACUTRON SUCTION REGULATOR DRAWINGS Figure 10 NOTE SEE SECTION 8 0 FOR ITEM IDENTIFICATION AND ORDERING INFORMATION...

Страница 13: ...300 mm Hg 1 22 95 2003 6 Gauge ISO With O Ring 0 To 300 mm Hg 1 22 95 2403 7 Gauge USA With O Ring 0 To 150 mm Hg 1 22 95 2023 7 Gauge ISO With O Ring 0 To 150 mm Hg 1 22 95 2423 8 O Rings For Mode S...

Страница 14: ...14 9 0 ISO USA MODELS...

Страница 15: ...herewith and provided further that in no event shall the aggregate liability of Allied in connection with breach of any warranty or warranties exceed the purchase price paid for the product purchased...

Страница 16: ...unswick NJ Krytox is a Registered Trademark of E I DuPont de Nemours Co Inc Wilmington DE Vacutron and Chemetron are Registered Trademarks of Allied Healthcare Products Inc 2001 Allied Healthcare Prod...