Command reference
R&S
®
ZND
974
User Manual 1173.9557.02 ─ 61
[SENSe<Ch>:]CORRection:CONNection:DELete
<ConnectorName>
Deletes a user-defined connector type named
<ConnectorName>
.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<ConnectorName>
Name of the user-defined connectors, string parameter.
Example:
See
[SENSe<Ch>:]CORRection:CONNection
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>?
[<SfkSettingType>[, <Index>]]
Gets the settings of the active system error correction (or of the factory calibration, if no
channel calibration is active).
Suffix:
<Ch>
.
Number of the calibrated channel.
<Sfk>
Number of the system error correction.
Must be omitted or set to 1.
Query parameters:
<SfkSettingType>
ACAL | STARt | STOP | POINts | SPOWer | STYPe |
BANDwidth | PDLY | RATTenuation | TYPE | PORTs | SPORt |
THRoughs | TSTamp | LTSTamp | TVNA | MVNA | MTESt |
CKIT | FSMode
The requested setting.
If no <SfkSettingType> is specified, the values for STARt, STOP,
POINts, SPOWer and STYPe are returned
ACAL
1 for automatic calibrations, 0 for manual calibrations
STARt
Start frequency (or CW frequency, if no frequency sweep is
active)
STOP
Stop frequency (or CW frequency, if no frequency sweep is
active)
POINts
Number of points
SPOWer
Source power (or stop power, for power sweeps)
STYPe
Sweep type or grid (LIN, LOG, SEGM)
BANDwidth
Measurement bandwidth
SCPI command reference
www.allice.de
Allice Messtechnik GmbH