
Command reference
R&S
®
ZND
954
User Manual 1173.9557.02 ─ 61
Parameters:
<Average>
AUTO | MANual
Manual operation:
[SENSe:]CORRection:COLLect:CHANnels:ALL
<Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean>
ON
–
calibrate all channels.
OFF
–
calibrate the active channel only.
Example:
[SENSe<Ch>:]CORRection:COLLect:SAVE:
Manual operation:
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
<Boolean>
Toggles the
M
ultiple
C
alibration
TYpes
mode that allows you to calibrate a subset of
the available channels using channel-specific ports and calibration types (see
ter 4.5.7, "Parallel calibration of multiple channels"
Enable the MCTYpes mode before defining the calibrations to be performed using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
. Disable the MCTYpes
mode after the calibrations were saved using
. The latter must be called only
once
, with <Ch>
referring to one of the channels to be calibrated (see the example below).
Parameters:
<Boolean>
SCPI command reference
www.allice.de
Allice Messtechnik GmbH