Command reference
R&S
®
ZND
809
User Manual 1173.9557.02 ─ 61
Manual operation:
See
CALCulate<Chn>:IAVerage[:STATe]
<Boolean>
Turns infinite averaging on or off.
The averaging mode can be selected using
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
CALCulate<Chn>:PHOLd
<HoldFunc>
Enables, disables, or restarts the max hold and the min hold functions.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HoldFunc>
MIN | MAX | OFF
MIN - Enable the min hold function.
MAX - enable the max hold function.
OFF - disable the max hold or min hold function.
*RST:
OFF
Example:
*RST; :CALC:PHOL MAX
Reset the instrument and enable the max hold function.
CALC:PHOL OFF; PHOL MAX
Restart max hold.
Manual operation:
See
CALCulate:PIPelining:DATA?
[<DutId>]
Queries the availability of channel results (to be queried, e.g., using
) in
single-sweep
mode.
The string result
'<true|false>,<channel nos>'
indicates:
●
false
, if the sweep is still running
●
true
, if the sweep has been completed
●
The comma-separated list of
channels nos
for which new measurement data
are available (since the last call of
CALC:PIP:DATA?
)
Channels measured in
continuous
mode are not considered, and not returned.
SCPI command reference
www.allice.de
Allice Messtechnik GmbH